Impact of secondary electron emission noise in SEM

Microscopy ◽  
2019 ◽  
Vol 68 (4) ◽  
pp. 279-288 ◽  
Author(s):  
Makoto Sakakibara ◽  
Makoto Suzuki ◽  
Kenji Tanimoto ◽  
Yasunari Sohda ◽  
Daisuke Bizen ◽  
...  

Abstract In semiconductor-device inspection using scanning electron microscopes (SEMs), the irradiation dose of the electron beam becomes lower because of increasing needs for higher throughput and lower damage to the samples. Therefore, it is necessary to form images using fewer primary electrons, making noise reduction of SEM images one of the main challenges. We have modeled the imaging process of SEMs, which consists of the generation of primary, secondary and tertiary electrons (PEs, SEs and TEs, respectively), and detection. Furthermore, a method to accurately evaluate the fluctuation in the number of SEs and TEs are proposed. We found that SEM-image noise can be minimized by directly detecting SEs generated in the sample, in which case the fluctuation in the number of SEs determines the image quality. The variance number of SEs emitted from a 500-eV PE irradiation onto a Si wafer is 1.9 times as large as the value derived assuming a Poisson process. A Monte-Carlo simulation result was used to explain the experimental results and predict that PE energy less than 1 keV suppresses the fluctuation in the number of SEs, and consequently, the SEM-image noise level. These findings provide a method for determining imaging conditions that improve the throughput of SEMs.

Author(s):  
Nelson C. Yew

There are two major areas where Dynamic Focusing techniques can be used to enhance SEM image qualify and micrograph information content. When used in conjunction with the final condenser lens in the electron-optical column, it increases the effective depth of focus associated with an inclined specimen in the SEM. When used with an ultra-high resolution recording cathode ray tube it provides exceptional corner-to-corner sharpness on the micrograph.Most commercial scanning electron microscopes use a tilted specimen positioned close to the final condenser lens with the secondary electron collector system located in the tilting direction to facilitate efficient electron collection. electron beam pivoting through the center of the principle plane of this lens to minimize distortion and off-axis aberration problems. As a result, the final electron beam is truly focused only along a portion of the spherical surface with its center located at the pivoting point.


1997 ◽  
Vol 3 (S2) ◽  
pp. 1193-1194
Author(s):  
Brendan J. Griffin

Most scanning electron microscopy is performed at low magnification; applications utilising the large depth of field nature of the SEM image rather than the high resolution aspect. Some environmental SEMs have a particular limitation in that the field of view is restricted by a pressure limiting aperture (PLA) at the beam entry point of the specimen chamber. With the original ElectroScan design, the E-3 model ESEM utilised a 500 urn aperture which gave a very limited field of view (∼550um diameter at a 10mm working distance [WD]). An increase of aperture size to ∼lmm provided an improved but still unsatisfactory field of view. The simplest option to increase the field of view in an ESEM was noted to be a movement of the pressure and field, limiting aperture back towards the scan coils1. This approach increased the field of view to ∼2mm, at a 10mm WD. A commercial low magnification device extended this concept and indicated the attainment of conventional fields of view.


Author(s):  
H. S Schmiβer ◽  
R. A. Swensson

Development of new signal detection systems for STEREOSCAN scanning electron microscopes has greatly increased the applicability of quantitative image analysis to pictures obtained with an s.e.m. While quantitative image analysis has proven to be a powerful technique to draw quantitative information from pictures obtained from light microscopes as well as various other sources of images, the field of scanninq electron microscopy has largely been uncovered in the past.This was mainly due to the particular appearance of “normal” s.e.m. pictures. Exactly those features which make s.e.m. pictures easy to interpret intuitively an aesthetically attractive i.e. the pseudo three dimensional appearance made it virtually impossible to treat these images with the well known methods of electronic image analysis.


2000 ◽  
Vol 6 (S2) ◽  
pp. 794-795
Author(s):  
B.L. Thiel ◽  
I.C. Bache ◽  
P. Smith

In low vacuum scanning electron microscopes, the primary beam is partially scattered by the gas present in the specimen chamber. The development of these microscopes, in particular the so-called ‘Environmental’ SEM, was initiated when it was realized that this scattering does not necessarily compromise the imaging capabilities of the instrument. Indeed, some modern commercial instruments are capable of better than 2 nanometer resolution at gas pressures of several torr. The accepted explanation for this is as follows: The mean-free-path of the high energy primary electrons is several millimeters in one torr of water vapour (for example). Because the actual pathlength of electrons travelling through the gas is only a few millimeters, most of them do not scatter at all. Those that do scatter are supposedly distributed over a relatively large area. Thus, the probe features a high-intensity central region surrounded by a slowly decaying low-intensity skirt. High resolution imaging is possible because the signal-to-(skirt)background ratio is high.


2003 ◽  
Vol 11 (6) ◽  
pp. 26-29 ◽  
Author(s):  
Michael D.G. Steigerwald

LEO's field emission scanning electron microscopes are all based an the “GEMINI” principle as shown in figure 1. In order to reduce aberrations and sensitivity to interfering stray-fields the electron optical column possesses a positively biased booster that shifts the energy of the primary electrons. The incident beam is focussed by a combination of a magnetic lens with an axial gap that avoids field leakage to the specimen and an electrostatic retarding lens formed by the beam booster together with the grounded pole piece cap. Shortly before the electrons hit the specimen they are decelerated down to the desired primary energy.


2015 ◽  
Vol 231 ◽  
pp. 139-144
Author(s):  
Anna Wassilkowska ◽  
Teresa Woźniakiewicz

In environmental monitoring and water protection one has to deal with various hydrated specimens, and therefore, there is a need for scanning electron microscopes which are suited for hydrated specimens. Wet specimens can be observed in a hydrated state using a commercial Peltier cooling holder for conventional microscopes. However, the optimization of the time of observation for the sample needs to be determined by carrying out experiments in advance for specifically determining the optimal observation time. The temperature-pressure conditions also require optimization to obtain good quality images of amoeba and fungi. Further studies are required to establish whether the same imaging conditions would be possible to replicate closely the observation on other biological specimens.


F1000Research ◽  
2021 ◽  
Vol 10 ◽  
pp. 875
Author(s):  
Khoa Van Pham

Background: Surface roughness is one of the most important characteristics of endodontic instruments, correlating to instrument fracture. The purpose of this study was to measure the surface roughness values of these instruments before and after resin root canal preparation without previous glide path preparation, with the new method. Data was obtained from field emission scanning electron microscopes (FE-SEM) combined with independent ImageJ software (NIH, Bethesda, MD, USA). Methods: A total of 20 simulated J-shape resin blocks with a radius of 4.5 mm, length of 16 mm, and angle of inflection of 600 were chosen and distributed into two equal groups. Each group was prepared by the WaveOne Gold Primary (Dentsply Sirona, Maillefer, Ballaigues, Switzerland) or the Reciproc Blue R25 (VDW, Munich, Germany) instruments, without glide path preparation. Special molds were used to confirm the same areas on the cutting blade at 3 mm and the instruments’ tips were scanned by FE-SEM, at different observed times. The parameters of Ra, Rq, and Rz in each sample were collected using the ImageJ software for analyses. The data was processed using the paired t-test with a significance level of 0.05. Results: Right after the first resin canal instrumentation, the surface roughness parameters of the two reciprocating investigated instruments were decreased. Conclusions: The FE-SEM images processed using the ImageJ software offered a trustworthy and suitable method for assessment of the NiTi endodontic file surface roughness.


1996 ◽  
Vol 4 (8) ◽  
pp. 18-19
Author(s):  
M.T. Postek ◽  
A.E. Vladar

Fully automated or semi-automated scanning electron microscopes (SEM) are now commonly used in semiconductor production and other forms of manufacturing. The industry requires that an automated instrument must be routinely capable of 5 nm resolution (or better) at 1.0 kV accelerating voltage for the measurement of nominal 0.25-0.35 micrometer semiconductor critical dimensions. Testing and proving that the instrument is performing at this level on a day-by-day basis is an industry need and concern which has been the object of a study at IMIST. The fundamentals and results are discussed in this paper.


Author(s):  
C. Colliex ◽  
P. Trebbia

The physical foundations for the use of electron energy loss spectroscopy towards analytical purposes, seem now rather well established and have been extensively discussed through recent publications. In this brief review we intend only to mention most recent developments in this field, which became available to our knowledge. We derive also some lines of discussion to define more clearly the limits of this analytical technique in materials science problems.The spectral information carried in both low ( 0<ΔE<100eV ) and high ( >100eV ) energy regions of the loss spectrum, is capable to provide quantitative results. Spectrometers have therefore been designed to work with all kinds of electron microscopes and to cover large energy ranges for the detection of inelastically scattered electrons (for instance the L-edge of molybdenum at 2500eV has been measured by van Zuylen with primary electrons of 80 kV). It is rather easy to fix a post-specimen magnetic optics on a STEM, but Crewe has recently underlined that great care should be devoted to optimize the collecting power and the energy resolution of the whole system.


Sign in / Sign up

Export Citation Format

Share Document