Modified flame ionization detector for supercritical fluid chromatography

1984 ◽  
Vol 56 (4) ◽  
pp. 831-832 ◽  
Author(s):  
Michael G. Rawdon
2004 ◽  
Vol 82 (3) ◽  
pp. 479-482 ◽  
Author(s):  
Kevin B Thurbide ◽  
Shawn Gilbert

The design and characterization of a flame ionization detector (FID) that is compatible with the high flow rates of packed-column supercritical fluid chromatography (pSFC) is presented. The modified FID uses a large quartz burner (4.0 mm i.d.) in place of the conventional burner (0.4 mm i.d.) and operates a stable flame in the presence of gas flows tested up to nearly 3000 mL/min. This modification allows the full flow of effluent from a pSFC column (i.e., splitless mode) to be directed into the FID. Results demonstrate that the modified FID yields response that is within a factor of 3 to that of a commercial FID. The attributes of the modified detector indicate that this approach could be a reliable and relatively simple means of achieving a splitless pSFC-FID system.Key words: supercritical fluid chromatography, packed column, splitless, flame ionization detector.