scholarly journals Development and Validation of Liquid Chromatography-Tandem Mass Spectrometry Method for Simple Analysis of Sumatriptan and its Application in Bioequivalence Study

2020 ◽  
Vol 13 (2) ◽  
pp. 21
Author(s):  
Wisut Wichitnithad ◽  
Siriwan Nantaphol ◽  
Petploy Vicheantawatchai ◽  
Thanyaporn Kiatkumjorn ◽  
Wachirasak Wangkangwan ◽  
...  

This work demonstrated a sensitive, selective, and simple liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for quantitation of sumatriptan in human plasma samples. Terazosin was used as an internal standard to minimize the variability during sample processing and detection. Sample cleanup prior to chromatographic analysis was accomplished by liquid-liquid extraction (LLE) with tert-butyl methyl ether (t-BME). The separation was performed on a reversed-phase Symmetry® C18 column (150 × 4.6 mm i.d., 5 µm) under a gradient mode, using a 0.2% formic acid aqueous solution and acetonitrile at a flow rate of 0.5 mL/min. Sumatriptan (m/z 296.26→251.05) and terazosin (m/z 388.10→290.25) were quantified using a triple quadrupole mass spectrometer, operating in the multiple reaction monitoring (MRM) under the positive ion mode. The method was fully validated following US-FDA and EMA guidelines. The LC-MS/MS assay had a calibration range of 0.5–50.0 ng/mL. The assay was precise and accurate with a between-run precision of <9.51%, and between-run accuracy between −7.27 to 8.30%. The developed method was subsequently applied in the determination of plasma concentration-time profile of a sumatriptan 50-mg tablet following oral administration in healthy volunteers.

Molecules ◽  
2020 ◽  
Vol 25 (7) ◽  
pp. 1600 ◽  
Author(s):  
Essam Ezzeldin ◽  
Muzaffar Iqbal ◽  
Yousif A. Asiri ◽  
Azza A Ali ◽  
Prawez Alam ◽  
...  

Baricitinib, is a selective and reversible Janus kinase inhibitor, is commonly used to treat adult patients with moderately to severely active rheumatoid arthritis (RA). A fast, reproducible and sensitive method of liquid chromatography-tandem mass spectrometry (LC-MS/MS) for the quantification of baricitinib in rat plasma has been developed. Irbersartan was used as the internal standard (IS). Baracitinib and IS were extracted from plasma by liquid–liquid extraction using a mixture of n-hexane and dichloromethane (1:1) as extracting agent. Chromatographic separation was performed using Acquity UPLC HILIC BEH 1.7 µm 2.1 × 50 mm column with the mobile phase consisting of 0.1% formic acid in acetonitrile and 20 mM ammonium acetate (pH 3) (97:3). The electrospray ionization in the positive-mode was used for sample ionization in the multiple reaction monitoring mode. Baricitinib and the IS were quantified using precursor-to-production transitions of m/z 372.15 > 251.24 and 429.69 > 207.35 for baricitinib and IS, respectively. The method was validated according to the recent FDA and EMA guidelines for bioanalytical method validation. The lower limit of quantification was 0.2 ng/mL, whereas the intra-day and inter-day accuracies of quality control (QCs) samples were ranged between 85.31% to 89.97% and 87.50% to 88.33%, respectively. Linearity, recovery, precision, and stability parameters were found to be within the acceptable range. The method was applied successfully applied in pilot pharmacokinetic studies.


Author(s):  
BG Keevil ◽  
SJ McCann ◽  
DP Cooper ◽  
MR Morris

Background: The immunosuppressive drug tacrolimus has complex and unpredictable pharmacokinetics, therefore regular monitoring is required in patients receiving tacrolimus therapy. We have developed a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for measuring tacrolimus concentrations in whole blood and have compared it with a microparticle enzyme immunoassay. Methods: For the LC-MS/MS assay, samples were prepared in a 96-deep well microtitre plate by adding 10 µL of blood to 40 µL of 0·1 mol/L zinc sulphate solution. Proteins were precipitated by adding 100 µL acetonitrile containing ascomycin internal standard. After vigorous mixing and centrifugation, 20 µL of the supernatant was injected into the LC-MS/MS system. A C18 cartridge (3 mm × 4 mm) was eluted with a step gradient of 50% to 100% methanol containing 2 mmol/L ammonium acetate and 0·1% (v/v) formic acid, at 0·6 mL/min. The column was maintained at 55°C. Results: The retention times were 0·98 min for ascomycin and 0·98 min for tacrolimus. Cycle time was 2·5 min, injection to injection. The analytes were monitored using a Quattro micro tandem mass spectrometer operated in multiple reaction monitoring mode using the following transitions: m/z821 > 768 (tacrolimus) and m/z809 > 756 (ascomycin). The limit of quantitation was 0·5 µg/L and the assay was linear up to 30 µg/L. Precision of the method, over the concentration range 2·5-15·0 µg/L, was < 7% within-batch and < 6% between-batch. Total time to analyse 24 samples including result generation was 90 min. Conclusion: We conclude that the LC-MS/MS method is quick, precise and robust and will provide a fast turn around of results for the transplant physician.


Author(s):  
DEEPAN T ◽  
BASAVESWARA RAO MV ◽  
DHANARAJU MD

Objective: A validated liquid chromatography–tandem mass spectrometry (LC–MS/MS) method was developed for canagliflozin in human plasma along with stability studies. Methods: The chromatographic separation of canagliflozin was performed on Zorbax XDB phenyl (75 × 4.6 mm, 3.5 mm) using methanol:acetate buffer (80:20 v/v) at a flow rate of 1.0 ml/min. The LC–MS/MS system consists of API 4000 triple quadrupole mass spectrometer equipped with turbospray ionization and an AS8020 automatic sample injector. Results: The retention time of canagliflozin was 1.15 min and total runtime was 2 min. The multiple reaction monitoring was 462.5/267.1 (m/z) for canagliflozin and 466.4/267.2 (m/z) for internal standard (canagliflozin D4), respectively. The method was linear over the range of 10–7505 ng/ml. The calculated slope ranged from 0.0451 to 0.0502 and intercepts from 0.0102 to 0.0456 with coefficients of the determination of 0.9970. The overall mean recovery of internal standard and canagliflozin was 76.66 and 79.77, respectively. Conclusion: The method was successfully validated and it was found to be within the limits for accuracy, precision, and linearity and it is stable under analytical conditions used.


2020 ◽  
Vol 2020 ◽  
pp. 1-8
Author(s):  
Yan Du ◽  
Hongliang Su ◽  
Jie Cao ◽  
Zhiwen Wei ◽  
Yujin Wang ◽  
...  

Male Sprague-Dawley rats (n=18) were randomly divided into three groups: a saline group (20 mL/kg by gavage), a ketamine (KET) group (100 mg/kg by gavage), and a KET (the same routes and doses) combined with levo-tetrahydropalmatine (l-THP; 40 mg/kg by gavage) group (n=6). Blood samples were acquired at different time points after drug administration. A simple and sensitive ultraperformance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) method was established to determine the concentrations of KET and its metabolite, norketamine (NK), in rat plasma. Chromatographic separation was achieved using a BEH C18 column (2.1 mm×50 mm, 1.7 μm) with chlorpheniramine maleate (Chlor-Trimeton) as an internal standard (IS). The initial mobile phase consisted of acetonitrile–water with 0.1% methanoic acid (80 : 20, v/v). The multiple reaction monitoring (MRM) modes of m/z 238.1→m/z 179.1 for KET, m/z 224.1→m/z 207.1 for NK, and m/z 275→m/z 230 for Chlor-Trimeton (IS) were utilized to conduct a quantitative analysis. Calibration curves of KET and NK in rat plasma demonstrated good linearity in the range of 2.5–500 ng/mL (r>0.9994), and the lower limit of quantification (LLOQ) was 2.5 ng/mL for both. Moreover, the intra- and interday precision relative standard deviation (RSD) of KET and NK were less than 4.31% and 6.53%, respectively. The accuracies (relative error) of KET and NK were below -1.41% and -6.07%, respectively. The extraction recoveries of KET and NK were more than 81.23±3.45% and 80.42±4.57%, respectively. This sensitive, rapid, and selective UPLC-MS/MS method was successfully applied to study the pharmacokinetic effects of l-THP on KET after gastric gavage. The results demonstrated that l-THP could increase the bioavailability of KET and promote the metabolism of KET. The results showed that l-THP has pharmacokinetics effects on KET in rat plasma.


Author(s):  
David J Marshall ◽  
Joanne E Adaway ◽  
Brian G Keevil

Background Analysis of citrate and oxalate in a 24-h urine sample is important in the screening and monitoring of patients with nephrolithiasis. To streamline the analytical process, it was decided to combine oxalate and citrate and analyse them simultaneously in the same assay. Objective A highly sensitive and specific assay for analysis of urine citrate and oxalate was developed using liquid chromatography-tandem mass spectrometry (LC-MS/MS) with a simple weak anion exchange solid phase extraction (WAX SPE) clean-up procedure. Method Premixed calibrator/acidified urine (50  µL) was combined with mixed internal standard (13C2 oxalate/citrate-d4) and 5% v/v formic acid in water and passed through a Waters WAX SPE plate. After clean-up steps, the plate was eluted with 5% NH3 in methanol, the eluent was dried down and re-constituted with 100  µL distilled water. Separation was then performed on an HSS T3 2.1 × 50 mm column (Waters, Manchester, UK), flow rate of 0.5 mL/min using a gradient of aqueous and organic mobile phases. We detected multiple reaction monitoring transitions m/z citrate 191.1>110.9, citrate IS 195.1>112.9, oxalate 88.9>60.85, oxalate IS 90.9>61.9 using a Waters TQD in electrospray-negative mode. Results Oxalate and 13C2 oxalate were eluted at 0.29 min; citrate and citrate-d4 were eluted at 0.52 min. Mean recovery was 100% for oxalate and 103% for citrate; lower limit of quantification of oxalate was 60  µmol/L and 50  µmol/L for citrate. Oxalate was linear up to 1388  µmol/L; citrate was linear up to 4762.5  µmol/L. Oxalate was found to be affected by ion suppression (matrix effect: −23 to +65%) but was compensated for by the internal standard used in all cases. The coefficient of variation of the assay in urine for oxalate was <7% for oxalate and 5% for citrate. Discussion We have developed a rapid assay for LC-MS/MS measurement of urinary oxalate and citrate in a routine clinical laboratory. It is simple, reproducible and easy to perform.


2010 ◽  
Vol 61 (4) ◽  
pp. 381-388 ◽  
Author(s):  
Robert Roškar ◽  
Marija Dolenc

Determination of Benzodiazepines in Urine Via Benzophenone Derivatives Using Liquid Chromatography-Tandem Mass SpectrometryThe aim of this study was to validate a new method for determining benzodiazepines in urine via their benzophenone derivatives, based on liquid chromatography-tandem mass spectrometry (LC-MS/MS). Selected benzodiazepines were analysed after acid hydrolysis of urine and extraction by ethyl acetate in the presence of an internal standard. Samples were analysed using electrospray ionization LC-MS/MS in a multiple reaction monitoring mode. The chromatographic run time on a reversed phase C18 analytical column was set for 9 min. This method was validated in 21 patients receiving methadone. Benzodiazepines intake was established in two out of three patients. LC-MS/MS results were also compared with the rapid immunoassay and the methods showed good agreement. However, in three cases benzodiazepines were detected by LC-MS/MS, but not by the immunoassay. The sensitivity of the developed LC-MS/MS method is comparable to or even higher than of previously reported methods, which makes it suitable as a confirmatory method.


2014 ◽  
Vol 2014 ◽  
pp. 1-7 ◽  
Author(s):  
Yunliang Zheng ◽  
Xingjiang Hu ◽  
Jian Liu ◽  
Guolan Wu ◽  
Huili Zhou ◽  
...  

A highly sensitive, simple, and rapid liquid chromatography tandem mass spectrometry method to simultaneously determine blonanserin and blonanserin C in human plasma with AD-5332 as internal standard (IS) was established. A simple direct protein precipitation method was used for the sample pretreatment, and chromatographic separation was performed on a Waters XBridge C8(4.6×150 mm, 3.5 μm) column. The mobile phase consists of a mixture of 10 mM ammonium formate and 0.1% formic acid in water (A) and 0.1% formic acid in methanol (B). To quantify blonanserin, blonanserin C, and IS, multiple reaction monitoring (MRM) was performed in positive ESI mode. The calibration curve was linear in the concentration range of 0.012–5.78 ng·mL−1for blonanserin and 0.023–11.57 ng·mL−1for blonanserin C (r2>0.9990). The intra- and interday precision of three quality control (QC) levels in plasma were less than 7.5%. Finally, the current simple, sensitive, and accurate LC-MS/MS method was successfully applied to investigate the pharmacokinetics of blonanserin and blonanserin C in healthy Chinese volunteers.


Author(s):  
Ana Carolina Furiozo Arantes ◽  
Kelly Francisco da Cunha ◽  
Marilia Santoro Cardoso ◽  
Karina Diniz Oliveira ◽  
Jose Luiz Costa

Abstract Purpose We developed and validated a method for quantitative analysis of 50 psychoactive substances and metabolites (antidepressants, benzodiazepines and opioids) in oral fluid samples using simple liquid–liquid extraction procedure followed by liquid chromatography–tandem mass spectrometry (LC–MS/MS). Method Oral fluid samples were collected using Quantisal™ device and extracted by liquid–liquid extraction with 1.0 mL of methyl tert-butyl ether and then analyzed using LC–MS/MS. Results The method attended method validation criteria, with limits of quantification as low as 0.5 and 1.0 ng/mL, and linearity between 0.5–50.0 ng/mL for antidepressants, 0.5–25.0 ng/mL for benzodiazepines and 1.0–50.0 ng/mL to opioids. During method validation, bias and imprecision values were not greater than 16 and 20%, respectively. Ionization suppression/enhancement bias results were not greater than 25%. No evidence of carryover was observed. Sample stability studies showed that almost all analytes were stable at 25 °C for 3 days and at 4 °C for 7 days. Freeze–thaw cycles stability showed that most antidepressants and opioids were stable under these conditions. Autosampler stability study showed that all analytes were stable for 24 h, except for nitrazepam and 7-aminoclonazepam. Thirty-eight authentic oral fluid samples were analyzed; 36.8% of the samples were positive for 2 drugs. Citalopram was the most common drug found, followed by venlafaxine. Conclusions The method was validated according to international recommendations for the 50 analytes, showing low limits of quantification, good imprecision and bias values, using simple liquid–liquid extraction, and was successfully applied to authentic oral fluid samples analysis.