scholarly journals Development of a Clinical Interface for a Novel Newborn Resuscitation Device: Human Factors Approach to Understanding Cognitive User Requirements

10.2196/12055 ◽  
2019 ◽  
Vol 6 (2) ◽  
pp. e12055 ◽  
Author(s):  
Laura Pickup ◽  
Alexandra Lang ◽  
Lara Shipley ◽  
Caroline Henry ◽  
James Carpenter ◽  
...  

Background A novel medical device has been developed to address an unmet need of standardizing and facilitating heart rate recording during neonatal resuscitation. In a time-critical emergency resuscitation, where failure can mean death of an infant, it is vital that clinicians are provided with information in a timely, precise, and clear manner to capacitate appropriate decision making. This new technology provides a hands-free, wireless heart rate monitoring solution that easily fits the clinical pathway and procedure for neonatal resuscitation. Objective This study aimed to understand the requirements of the interface design for a new device by using a human factors approach. This approach combined a traditional user-centered design approach with an applied cognitive task analysis to understand the tasks involved, the cognitive requirements, and the potential for error during a neonatal resuscitation scenario. Methods Fourteen clinical staff were involved in producing the final design requirements. Two pediatric doctors supported the development of a visual representation of the activities associated with neonatal resuscitation. This design was used to develop a scenario-based workshop. Two workshops were carried out in parallel and involved three pediatric doctors, three neonatal nurses, two advance neonatal practitioners, and four midwives. Both groups came together at the end to reflect on the findings from the separate sessions. Results The outputs of this study have provided a comprehensive description of information requirements during neonatal resuscitation and enabled product developers to understand the preferred requirements of the user interface design for the device. The study raised three key areas for the designers to consider, which had not previously been highlighted: (1) interface layout and information priority, as heart rate should be central and occupy two-thirds of the screen; (2) size and portability, to enable positioning of the product local to the baby’s head and allow visibility from all angles; and (3) auditory feedback, to support visual information on heart rate rhythm and reliability of the trace with an early alert for intervention while avoiding parental distress. Conclusions This study demonstrates the application of human factors and the applied cognitive task analysis method, which identified previously unidentified user requirements. This methodology provides a useful approach to aid development of the clinical interface for medical devices.

2018 ◽  
Author(s):  
Laura Pickup ◽  
Alexandra Lang ◽  
Lara Shipley ◽  
Caroline Henry ◽  
James Carpenter ◽  
...  

BACKGROUND A novel medical device has been developed to address an unmet need in standardising and facilitating heart rate recording during neonatal resuscitation. In a time critical emergency resuscitation, where failure can mean death of an infant, it is vital that clinicians are provided with information in a timely, precise and clear manner to capacitate appropriate decision making. This new technology provides a hands free, wireless heart rate monitoring solution that easily fits the clinical pathway and procedure for neonatal resuscutation. To understand the requirements of the interface design for this new device, a human factors approach was implemented. This combined a traditional user-centred design approach with an Applied Cognititive Task Analysis (ACTA) to understand the tasks involved, the cognitive requirements and the potential for error during a neonatal resusciation scenario. OBJECTIVE 1. To understand the cognitive requirements of clinicians for a novel medical device to facilitate neonatal resuscitation; 2. To apply a human factors approach and a traditional user-centred design approach to provide a device interface specification. METHODS Fourteen clinical staff were involved in producing the final design requirements. Two paediatric doctors supported the development of a visual representation of the activities associated with neonatal resucitation. This was used to develop a scenario based workshop. Two workshops were carried out in parallel and involved three paediatric doctors, three neonatal nurses, two advance neonatal practitioners and four midwives. Both groups came together at the end to reflect on the findings which emerged during the separate sessions. RESULTS The outputs of this study have provided a comprehensive description of information requirements during neonatal resuscitation, and enabled product developers to understand the core and preferred requirements of the user interface design for the device. The study raised three key areas for the designers to consider, which had not previously been highlighted. These related to interface layout and information priority, size and portability of the device and auditory feedback. CONCLUSIONS This study demonstrates the value of the ACTA approach to inform the development of resuscitation devices, and more generally for medical device development.


2019 ◽  
Vol 7 ◽  
Author(s):  
Emily C. Zehnder ◽  
Brenda H. Y. Law ◽  
Georg M. Schmölzer

Author(s):  
Kim J. Vicente

Cognitive task analysis (CTA) is increasingly being used to effectively address a wide variety of human factors problems. However, different researchers are using significantly different methods. In many cases, a particular method is used solely by its originators. Therefore, there are significant issues that must be worked through before CTA becomes a widely accepted and easily transferable human factors tool. The objectives of this symposium are to: bring CTA to the attention of a wider audience; develop a better understanding of the differences and similarities between different CTA methods; and demonstrate the practical advantages of CTA.


2005 ◽  
Author(s):  
David E. Kancler ◽  
Megan E. Gorman ◽  
Mary McWesler ◽  
Scott Bachmann ◽  
Darryl Stimson ◽  
...  

Author(s):  
Laura Lin ◽  
Racquel Isla ◽  
Karine Doniz ◽  
Heather Harkness ◽  
Kim J. Vicente ◽  
...  

The hypothesis explored in this paper is that, by adopting human factors design principles, the use of medical equipment can be made safer and more efficient We have selected a commercially available patient-controlled analgesia (PCA) machine as a vehicle to test this hypothesis. A cognitive task analysis of PCA usage, combined with a set of human factors design principles, led to a redesigned PCA interface. An experimental evaluation was conducted, comparing this new interface with the existing interface. The results show that the new interface leads to significantly faster, less effortful, and more reliable performance. These findings have implications for improving the design of other medical equipment.


Sign in / Sign up

Export Citation Format

Share Document