scholarly journals Profound Trunnion Wear Resulting in Femoral Head-Neck Dissociation in Total Hip Arthroplasty

2018 ◽  
Vol 2018 ◽  
pp. 1-4 ◽  
Author(s):  
Saif Shamshoon ◽  
Patrick Thornley ◽  
Justin de Beer

We describe a case of aseptic failure with profound femoral stem trunnion wear and femoral head dissociation nine years after initial primary total hip arthroplasty (THA) with the Stryker Accolade total hip arthroplasty system. Current guidelines for postoperative care and follow-up after THAs as potential intervention points for early detection of prosthetic joint failure are also reviewed.

2017 ◽  
Vol 99-B (1) ◽  
pp. 29-36 ◽  
Author(s):  
I. De Martino ◽  
V. De Santis ◽  
R. D’ Apolito ◽  
P. K. Sculco ◽  
M. B. Cross ◽  
...  

2020 ◽  
Vol 62 (1) ◽  
Author(s):  
Đắc Việt Mai ◽  
Thu Thủy Nguyễn

This study aimed to report the outcome of total hip arthroplasty (THA) with an extensive hydroxyapatite (HA) coating for the fixation of a tapered femoral stem (Corail) in patients aged 60 or younger than sixty years with stage IV, V và VI osteonecrosis of femoral head. Subject and method: Descriptive prospective research of ninety osteonecrotic hips in ninety patients were available for clinical and radiographic analyses at minimum follow-up of 5 years. Results and Conclusion: The mean Harris hip score improved from 43.74 ± 9,25points preoperatively to 96.67 ± 3.82points at final follow-up. Seventy-four (100%) hips demonstrated stable bone ingrowth. No hips showed acetabular or femoral osteolysis radiolucency and loosening, or required revision for aseptic loosening. We believe that cementless THA with a Corail stem is a promising procedure for patients with osteonecrosis of the femoral head.


Medicina ◽  
2021 ◽  
Vol 57 (11) ◽  
pp. 1188
Author(s):  
Maximilian G. Hanslmeier ◽  
Michael W. Maier ◽  
Manuel Feisst ◽  
Nicholas A. Beckmann

Background and Objectives: Femoral head and liner exchange is an established treatment for polyethylene wear but has had a more limited role in the treatment of other conditions including dislocation, because of concerns about an increased postoperative dislocation rate. Some authors have considered dislocation associated with polyethylene wear to be a contraindication for this procedure. Materials and Methods: Our retrospective review evaluated the outcome of head and liner exchange in a small consecutively operated heterogeneous cohort of 20 patients who presented with dislocation unrelated to trauma, component malposition or component loosening. Of this group, 12 had prior primary total hip arthroplasty, and 8 had prior revision total hip arthroplasty, and included 4 patients with prior revision for dislocation. Mean follow-up was 6 ± 3.5 years (range 1–145 months). Results: Kaplan–Meier analysis revealed a revision-free implant survival from any cause of 80% (confidence interval 95%:64.3–99.6%) at 5 years after head and liner exchange (index surgery). At final follow-up, 83.3% of patients (n = 10) with prior primary total hip arthroplasty and 62.5% of patients (n = 5) with prior revision total hip arthroplasty, had not required subsequent revision for any cause. None (0%) of the primary total hip arthroplasty group and 3 (38%) of the revision arthroplasty group had required revision for further dislocation. Of the eight revision arthroplasty patients, four had a prior revision for dislocation and three of these four patients required further revision for dislocation after index surgery. The fourth patient had no dislocation after index surgery. One additional patient who had prior revision surgery for femoral component fracture suffered dislocation after index surgery, but was successfully treated with closed reduction. Conclusions: In our study population, femoral head and liner exchange was an effective treatment option for patients with prior primary total hip arthroplasty and also for a highly select group of revision total arthroplasty patients with no prior history of dislocation. Femoral head and liner exchange does not appear to be a viable treatment option for patients who have had revision total arthroplasty after prior dislocations.


2020 ◽  
pp. 112070002096964
Author(s):  
Kirill Gromov ◽  
Nanna H Sillesen ◽  
Thomas Kallemose ◽  
Henrik Husted ◽  
Henrik Malchau ◽  
...  

Background: Introduction of new implants should be monitored closely to capture any signs of compromising patient safety. Vitamin E infused highly-crosslinked polyethylene liners (VEPE) offer the potential for reduced wear. Highwall liners have been hypothesised to result in increased wear and potential liner fractures. The aim of this study was to determine the 3–7-year follow-up of highwall VEPE for primary total hip arthroplasty (THA), focusing on liner-related complications. Methods: We included 1221 consecutive THA operations from July 2010 to May 2014 with minimum follow-up of 3 (3.1–6.8) years Data collected included demographics, implant data, complications, reoperations, and deaths. Data were cross-referenced with the Danish Hip Arthroplasty Registry in order to ensure validity and completeness. Acetabular shell position was measured using Martell Hip Analysis Suite in a subgroup of 931 THAs. Results: Cumulative stem revision and shell revision at 3-year follow-up was 3.4% and 0.4% respectively. There were no revisions due to liner failure. Reason for revision included 11 dislocations, 15 soft-tissue revisions for infection, 44 stem revisions of which 34 were periprosthetic fractures and 13 shell revisions of which 6 were combined shell and stem revisions. Conclusion: Early follow-up of VEPE liners for primary THA have not shown any revisions associated with liner failure. Continued monitoring of new materials are necessary to capture any signs of compromised patient safety.