scholarly journals False Positive Transit Time Flowmetry Graft Failure in Multivessel Coronary Spasm following Off-Pump Coronary Artery Bypass Grafting

2017 ◽  
Vol 2017 ◽  
pp. 1-3
Author(s):  
George Kassimis ◽  
George Krasopoulos

Intraoperative Transit Time Flowmetry is currently recommended to assess graft patency during coronary artery bypass grafting (CABG), especially in presence of haemodynamic instability or inability to wean the patient from cardiopulmonary bypass, new regional wall motion abnormalities, or significant ventricular arrhythmias. The VeriQ™ system is one of the currently available systems, which detects imperfections that may be corrected by graft revision. In this case report, multivessel coronary spasm (CS) post-CABG interferes with these intraoperative parameters misleading initially into false results. Cardiac surgeons should bear in mind the limit of VeriQ in distinguishing between graft failure and CS. Angiography may be considered in patients with decreased graft flow despite revision of anastomosis and vasodilatory treatment for the definitive diagnosis.

JAMA Surgery ◽  
2014 ◽  
Vol 149 (11) ◽  
pp. 1182 ◽  
Author(s):  
Jacquelyn Quin ◽  
John Lucke ◽  
Brack Hattler ◽  
Sandeep Gupta ◽  
Janet Baltz ◽  
...  

2005 ◽  
Vol 79 (3) ◽  
pp. 854-857 ◽  
Author(s):  
Derek K.H. Leong ◽  
Venkataraman Ashok ◽  
Arulkumaran Nishkantha ◽  
Yue Hong Shan ◽  
Eugene K.W. Sim

Sign in / Sign up

Export Citation Format

Share Document