scholarly journals Surface and Buildup Region Dose Measurements with Markus Parallel-Plate Ionization Chamber, GafChromic EBT3 Film, and MOSFET Detector for High-Energy Photon Beams

2016 ◽  
Vol 2016 ◽  
pp. 1-10 ◽  
Author(s):  
Ugur Akbas ◽  
Nazmiye Donmez Kesen ◽  
Canan Koksal ◽  
Hatice Bilge

The aim of the study was to investigate surface and buildup region doses for 6 MV and 15 MV photon beams using a Markus parallel-plate ionization chamber, GafChromic EBT3 film, and MOSFET detector for different field sizes and beam angles. The measurements were made in a water equivalent solid phantom at the surface and in the buildup region of the 6 MV and 15 MV photon beams at 100 cm source-detector distance for 5 × 5, 10 × 10, and 20 × 20 cm2field sizes and 0°, 30°, 60°, and 80° beam angles. The surface doses using 6 MV photon beams for 10 × 10 cm2field size were found to be 20.3%, 18.8%, and 25.5% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface doses using 15 MV photon beams for 10 × 10 cm2field size were found to be 14.9%, 13.4%, and 16.4% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface dose increased with field size for all dosimeters. As the angle of the incident radiation beam became more oblique, the surface dose increased. The effective measurement depths of dosimeters vary; thus, the results of the measurements could be different. This issue can lead to mistakes at surface and buildup dosimetry and must be taken into account.

2008 ◽  
Vol 35 (2) ◽  
pp. 464-472 ◽  
Author(s):  
Giorgia Nicolini ◽  
Antonella Fogliata ◽  
Eugenio Vanetti ◽  
Alessandro Clivio ◽  
Daniel Vetterli ◽  
...  

2006 ◽  
Vol 49 (spe) ◽  
pp. 17-23 ◽  
Author(s):  
Carlos de Austerlitz ◽  
Viviane Souza ◽  
Heldio Pereira Villar ◽  
Aloisio Cordilha

The performance of four X-ray qualities generated in a Pantak X-ray machine operating at 30-100 kV was determined with a parallel-plate ionization chamber and a Fricke dosimeter. X-ray qualities used were those recommended by Deutsch Internationale Normung DIN 6809 and dose measurements were carried out with Plexiglas® simulators. Results have shown that the Fricke dosimeter can be used not only for soft X-ray dosimetry, but also for the maintenance of low-energy measuring systems' calibration factor.


2022 ◽  
Vol 12 (2) ◽  
pp. 600
Author(s):  
Serenella Russo ◽  
Silvia Bettarini ◽  
Barbara Grilli Leonulli ◽  
Marco Esposito ◽  
Paolo Alpi ◽  
...  

High-energy small electron beams, generated by linear accelerators, are used for radiotherapy of localized superficial tumours. The aim of the present study is to assess the dosimetric performance under small radiation therapy electron beams of the novel PTW microSilicon detector compared to other available dosimeters. Relative dose measurements of circular fields with 20, 30, 40, and 50 mm aperture diameters were performed for electron beams generated by an Elekta Synergy linac, with energy between 4 and 12 MeV. Percentage depth dose, transverse profiles, and output factors, normalized to the 10 × 10 cm2 reference field, were measured. All dosimetric data were collected in a PTW MP3 motorized water phantom, at SSD of 100 cm, by using the novel PTW microSilicon detector. The PTW diode E and the PTW microDiamond were also used in all beam apertures for benchmarking. Data for the biggest field size were also measured by the PTW Advanced Markus ionization chamber. Measurements performed by the microSilicon are in good agreement with the reference values for all the tubular applicators and beam energies within the stated uncertainties. This confirms the reliability of the microSilicon detector for relative dosimetry of small radiation therapy electron beams collimated by circular applicators.


Author(s):  
Tu Vu Ngoc

Purpose: Compare percent depth dose (PDD) and off-center ratio (OCR) measured by the CC13 ionization chamber and the RAZOR silicon diode in small photon beams. Method and Materials: Some dosimetric characteristics, such as PDD, OCR, penumbra and radiation field size, were considered in this study for 2x2, 3x3, and 4x4 cm2 field sizes. We used the CC13 ionization chamber and the RAZOR silicon diode to measure dose distribution with depth along the axis and off-center of the beam. From the results obtained, the team investigated the differences in radiation parameters measured by the two types of probes above. Results: There are significant differences in the radiation parameters investigated for the CC13 ionization chamber and the RAZOR silicon diode, especially the width of penumbra. For PDD curves, the difference is less than 5% from dmax to 30 cm, however the difference becomes greater in the build-up region, which reaches to 33% at the water phatom surface. The width of penumbra measured by CC13 is always larger than that of RAZOR, the ratio of the penumbra width between two detectors is 1.8 and 1.3 for energies of 6 MV and 15 MV, respectively. Conclusion: The RAZOR silicon diode has better dose response than the CC13 ionization chamber for measuring the PDD and the OCR in small photon beams.


2010 ◽  
Vol 10 (2) ◽  
pp. 77-83 ◽  
Author(s):  
ML Wilson ◽  
WP Colley ◽  
AW Beavis

AbstractCarbon fibre couch inserts are widely used in external beam radiotherapy to provide rigid and lightweight patient support. Carbon fibre is often perceived to be essentially radiotranslucent implying that it does not interfere with the radiation beam. However, there is evidence in the literature which suggests that this perception may not be appropriate, particularly at oblique angles of incidence. Furthermore, there is evidence indicating that the use of carbon fibre significantly reduces the skin sparing effect. In this study, the radiation attenuation and surface dose enhancement characteristics of the carbon fibre insert for the Varian ExactTM couch have been investigated. It was found that attenuation increased significantly with increasing angle of incidence, resulting in in-phantom dose reductions of up to 6% at 6 MV and 4% at 15 MV. It has been shown that it is possible to model couch attenuation on a commercial treatment planning system (Elekta CMS XiO) by including the carbon fibre insert in the planning computed tomography (CT) dataset. Finally, the carbon fibre insert was found to significantly increase skin dose to the patient. The skin dose was approximately three times as large when the couch insert was added to 6 and 15 MV photon beams. However, even with this substantial increase it is highly unlikely that the skin tolerance dose will be exceeded.