scholarly journals Metallosis after Exchange of the Femoral Head and Liner following Ceramic Acetabular Liner Dissociation in Total Hip Arthroplasty with a Modular Layered Acetabular Component

2016 ◽  
Vol 2016 ◽  
pp. 1-5
Author(s):  
Tomoya Takasago ◽  
Tomohiro Goto ◽  
Keizo Wada ◽  
Daisuke Hamada ◽  
Toshiyuki Iwame ◽  
...  

The type of bearing material that should be used in revision surgery after the failure of ceramic-on-ceramic total hip arthroplasty (THA) remains controversial. In the case of ceramic fracture, the residual ceramic particles can cause consequent metallosis when metal implants are used for revision THA. On the other hand, in the case of THA failure without ceramic fracture, revision THA with a metal femoral head provides satisfactory results. We report an unusual case of progressive osteolysis due to metallosis that developed after revision THA for ceramic liner dissociation without a liner fracture performed using a metal femoral head and polyethylene liner. The residual metal debris and abnormal pumping motion of the polyethylene liner due to the breakage of the locking system or the aspherical metal shell being abraded by the ceramic head seemed to be the cause of the progressive osteolysis.

2019 ◽  
Vol 101 (5) ◽  
pp. 421-428 ◽  
Author(s):  
Stephen M. Petis ◽  
Bernd Kubista ◽  
Robert U. Hartzler ◽  
Matthew P. Abdel ◽  
Daniel J. Berry

2014 ◽  
Vol 5 (2) ◽  
pp. 23-26 ◽  
Author(s):  
Joseph Schirmers, MD ◽  
Ryan Horazdovsky, MD ◽  
Scott Marston, MD

Introduced in 1974 by Bousquet, the dual-mobility bearing for use in total hip arthroplasty (THA) confers increased jump distance and improved overall stability relative to conventional THA designs [1-3]. The dual-mobility bearing incorporates a relatively small (22-28mm) metal or ceramic femoral head press fit into a larger polyethylene liner which articulates with the acetabular component. Dissociation of the femoral head from the polyethylene liner (intraprosthetic dislocation) is a known late complication thought to be related to polyethylene liner wear and has been previously reported [2-7]. In a consecutive series of 384 primary THAs employing Bousquet’s original design, there were 14 intraprosthetic dislocations over 15 years (3.6%). The authors cited polyethylene wear as causative and mean time to intraprosthetic dislocation was 8.9 years [2].


2017 ◽  
Vol 01 (02) ◽  
pp. 118-120 ◽  
Author(s):  
Sameer Oak ◽  
Salvatore Frangiamore ◽  
Carlos Rueda ◽  
Robert Molloy ◽  
Alison Klika

AbstractA 47-year-old man underwent a left ceramic-on-polyethylene total hip arthroplasty with a delta ceramic femoral head. At 20 months postoperatively, he presented with acute left groin pain after he stepped over a fishing rod while on a boat. A fracture of the femoral head without trauma was diagnosed upon presentation. Manufacturing practices with improvements in each successive generation have decreased historically high ceramic fracture rates considerably. Although femoral head fractures have proven to be extremely rare both in ceramic-on-polyethylene articulations and with the newest delta femoral head, it is important to remain vigilant in all situations involving ceramics.


2019 ◽  
Vol 4 (9) ◽  
pp. 541-547 ◽  
Author(s):  
William G. Blakeney ◽  
Jean-Alain Epinette ◽  
Pascal-André Vendittoli

Hip instability following total hip arthroplasty (THA) remains a major challenge and is one of the main causes of revision surgery. Dual mobility (DM) implants have been introduced to try to overcome this problem. The DM design consists of a small femoral head captive and mobile within a polyethylene liner. Numerous studies have shown that DM implants reduce the rate of dislocation compared to fixed-bearing inserts. Early designs for DM implants had problems with wear and intra-prosthetic dislocations, so their use was restricted to limited indications. The results of the latest generation of DM prostheses demonstrate that these problems have been overcome. Given the results of these studies presented in this review, surgeons may now consider DM THA for a wider patient selection. Cite this article: EFORT Open Rev 2019;4:541-547. DOI: 10.1302/2058-5241.4.180045


2017 ◽  
Vol 68 (5) ◽  
pp. 974-976
Author(s):  
Alexandru Patrascu ◽  
Liliana Savin ◽  
Dan Mihailescu ◽  
Victor Grigorescu ◽  
carmen Grierosu ◽  
...  

In recent years, there has been an increase in the number of studies on the etiology of femoral head necrosis. We retrospectively reviewed all patients diagnosed with aseptic necrosis of the femoral in the period of 2010-2015. We recorded a total of 230 cases diagnosed with aseptic necrosis of the femoral head, group was composed of 65.7% men and 34.3% women, risk factors identified was 19.13% (post-traumatic), 13.91% (glucocorticoids), 26.52% (alcohol), 3.47% (another cause) and in 36 95% of the cases no risk factors were found. The results of the study based on the type of surgery performed on the basis of stages of disease progression, 8 patients (3.48%) benefited from osteotomy, 28 patients (12.17%) benefited of bipolar hemiarthroplasty prosthesis and 188 patients (81.74%) benefited of total hip arthroplasty. Osteonecrosis of the femoral head is characteristic to young patients between the age of 30-50 years old. Predisposing factors, alcohol and corticosteroid therapy remains an important cause of the disease. Total hip arthroplasty remains the best option for the patients with osteonecrosis of the femoral head.


2021 ◽  
Vol 9 ◽  
pp. 205031212110147
Author(s):  
Nobuhiko Sumiyoshi ◽  
Kazuhiro Oinuma ◽  
Yoko Miura

Background: Adverse reactions to metal debris are significant complications after metal-on-metal total hip arthroplasty. Recently, late appearances of adverse reactions to metal debris and subsequent need for reoperations have been reported with small-diameter head metal-on-metal devices. We retrospectively investigated mid-term clinical outcomes of small-head metal-on-metal total hip arthroplasty. Methods: We reviewed 159 hips in 139 patients who had a small-head metal-on-metal total hip arthroplasty (M2a Taper; Biomet, Warsaw, IN) with a minimum 5-year follow-up and documented postoperative complications. Results: Focal osteolysis in either the femur or acetabulum was observed in 12 hips (7.5%, 44 months after surgery on average), with pseudotumor observed in 8 hips (5%, 120 months after surgery on average). Four hips (2.5%) had dislocations (84 months after surgery on average) and six hips (3.8%, 122 months after surgery on average) underwent reoperation. Conclusion: Small-head metal-on-metal total hip arthroplasty is associated with a high degree of complications at mid-term follow-up period. Considering this, we discourage the use of metal-on-metal total hip arthroplasty regardless of head size.