scholarly journals Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry

2015 ◽  
Vol 2015 ◽  
pp. 1-6 ◽  
Author(s):  
Jinhai Huang ◽  
Giacomo Savini ◽  
Chengfang Wang ◽  
Weicong Lu ◽  
Rongrong Gao ◽  
...  

Purpose.To assess the reliability and comparability of measuring central corneal thickness (CCT) and thinnest corneal thickness (TCT) using a new Scheimpflug-Placido analyzer (TMS-5, Japan) and ultrasound (US) pachymetry.Methods.Seventy-six healthy subjects were prospectively measured 3 times by 1 operator using the TMS-5, 3 additional consecutive scans were performed by a second operator, and ultrasound (US) pachymetry measurements were taken. The test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intraoperator repeatability and interoperator reproducibility. Agreement among the devices was assessed using Bland-Altman plots and 95% limits of agreement (LoA).Results.The intraoperators TRT and CoV were <19 μm and 2.0%, respectively. The interoperators TRT and CoV were <12 μm and 1.0%, respectively, and ICC was >0.90. The mean CCT and TCT measurements using the TMS-5 were 15.97 μm (95% LoA from −26.42 to −5.52 μm) and 20.32 μm (95% LoA from −30.67 to −9.97 μm) smaller, respectively, than those using US pachymetry.Conclusions.The TMS-5 shows good repeatability and reproducibility for measuring CCT and TCT in normal subjects but only moderate agreement with US pachymetry results. Caution is warranted before using these techniques interchangeably.

2015 ◽  
Vol 2015 ◽  
pp. 1-8 ◽  
Author(s):  
Ayong Yu ◽  
Weiqi Zhao ◽  
Giacomo Savini ◽  
Zixu Huang ◽  
Fangjun Bao ◽  
...  

Purpose.To assess the repeatability and reproducibility of central corneal thickness (CCT) measurements by corneal dynamic Scheimpflug analyzer Corvis ST in normal eyes and compare the agreement with Pentacam rotating Scheimpflug System and ultrasound pachymetry.Methods.84 right eyes underwent Corvis ST measurements performed by two operators. The test-retest repeatability (TRT), within-subject coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were used to evaluate the intraoperator repeatability and interoperator reproducibility. CCT measurements also were obtained from Pentacam and ultrasound pachymetry by the first operator. The agreement between the three devices was evaluated with 95% limits of agreement (LoA) and Bland-Altman plots.Results.Corvis ST showed high repeatability as indicated by TRT ≤ 13.0 μm, CoV < 0.9%, and ICC > 0.97. The interoperator reproducibility was also excellent. The CoV was <0.9%, and ICC was >0.97. Corvis ST showed significantly lower values than Pentacam and ultrasound pachymetry (P<0.001). The 95% LoA between Corvis ST and Pentacam or ultrasound pachymetry were −15.8 to 9.5 μm and −27.9 to 12.3 μm, respectively.Conclusions.Corvis ST showed excellent repeatability and interoperator reproducibility of CCT measurements in normal eyes. Corvis ST is interchangeable with Pentacam but not with ultrasound pachymetry.


2015 ◽  
Vol 2015 ◽  
pp. 1-6 ◽  
Author(s):  
Ayaka Iijima ◽  
Kimiya Shimizu ◽  
Hidenaga Kobashi ◽  
Aya Saito ◽  
Kazutaka Kamiya

Purpose. To assess the repeatability, reproducibility, and comparability of measurements of subjective and objective forward scattering in healthy subjects.Methods. We prospectively examined twenty eyes of 20 healthy volunteers (7 men and 13 women; ages, 28.4 ± 4.1 years). The logarithmic straylight value (log(s)) and the objective scattering index (OSI) were measured with a straylight meter (C-Quant) and a point-spread function meter (OQAS), respectively.Results. The 95% limits of agreement (LoA) between first and second measurements ranged from −0.211 to 0.207 for the C-Quant and from −0.302 to 0.477 for the OQAS. The intraclass correlation coefficients for the repeatability of the log(s) and OSI measurements were 0.815 and 0.926, respectively. The mean difference between examiners was −0.051 ± 0.133 (95% LoA; −0.311 to 0.209) for the C-Quant and 0.080 ± 0.307 (−0.522 to 0.682) for the OQAS. There was a modest, but significant, correlation between the log(s) and the OSI (Spearman correlation coefficientr=0.498,p=0.026).Conclusions. The C-Quant and the OQAS provide good repeatability and reproducibility, although the OQAS measurement provides a slightly higher ICC than the C-Quant measurement. The subjective forward scattering may be to some extent expressed in the objective forward scattering in healthy subjects.


2014 ◽  
Vol 2014 ◽  
pp. 1-6 ◽  
Author(s):  
Mehmet Kola ◽  
Hikmet Duran ◽  
Adem Turk ◽  
Suleyman Mollamehmetoglu ◽  
Ahmet Kalkisim ◽  
...  

Purpose. To assess the repeatability and reproducibility of ocular biometry and intraocular lens (IOL) power measurements obtained by ophthalmology residents using an AL-Scan device, a novel optical biometer.Methods. Two ophthalmology residents were instructed regarding the AL-Scan device. Both performed ocular biometry and IOL power measurements using AL-Scan, three times on each of 128 eyes, independently of one another. Corneal keratometry readings, horizontal iris width, central corneal thickness, anterior chamber depth, pupil size, and axial length values measured by both residents were recorded together with IOL power values calculated on the basis of four different IOL calculation formulas (SRK/T, Holladay, and HofferQ). Repeatability and reproducibility of the measurements obtained were analyzed using the intraclass correlation coefficient (ICC).Results. Repeatability (ICC, 0.872-0.999 for resident 1 versus 0.905-0.999 for resident 2) and reproducibility (ICC, 0.916-0.999) were high for all biometric measurements. Repeatability (ICC, 0.981-0.983 for resident 1 versus 0.995-0.996 for resident 2) and reproducibility were also high for all IOL power measurements (ICC, 0.996 for all).Conclusions. The AL-Scan device exhibits good repeatability and reproducibility in all biometric measurements and IOL power calculations, independent of the operator concerned.


2016 ◽  
Vol 2016 ◽  
pp. 1-5
Author(s):  
Jinhai Huang ◽  
Weicong Lu ◽  
Giacomo Savini ◽  
Hao Chen ◽  
Chengfang Wang ◽  
...  

Purpose.To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD).Methods.Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using pairedt-tests, Bland-Altman plots, and 95% limits of agreement (LoA).Results.The mean CCT, ACD, and AD measured by AL-Scan were538.59±27.37 μm,3.70±0.30 mm, and3.16±0.30 mm, respectively. The mean values obtained by the Visante OCT were536.14±26.61 μm for CCT,3.71±0.29 mm for ACD, and3.17±0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference =2.45±6.07 μm,P<0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively.Conclusions.Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic.


2021 ◽  
Vol 13 ◽  
pp. 251584142199563
Author(s):  
Mehmet Barış Üçer ◽  
Erdinç Bozkurt

Purpose: The purpose of this study was to compare and evaluate the agreement of central corneal thickness (CCT) values obtained with three different devices working according to optical principle in healthy eyes. Methods: 60 eyes of 60 individuals (30 men and 30 women) were enrolled in this study. CCT measurements performed with Scheimpflug–Placido topographer (Sirius), spectral-domain optical coherence tomography (RTVue) with an anterior segment module, and optical biometer (AL-Scan) were compared. Bland–Altman analysis was used to demonstrate agreement between methods. Results: The mean age was 30.07 ± 7.313 years (range, 18–47 years). The mean CCT values obtained by RTVue, Sirius, and AL-Scan were 518.25 ± 36.38 µm, 526.08 ± 36.33 µm, and 513.50 ± 39.09 µm, respectively. The mean differences in CCT were 7.83 ± 14.15 µm between Sirius and RTVue, 12.58 ± 11.87 µm between Sirius and AL-Scan, and 4.75 ± 4.50 µm between RTVue and AL-Scan. The mean CCT was statistically different among the three groups ( p < 0.05). All three modalities of CCT measurements correlated closely with each other, with Pearson’s correlation coefficients ranging from 0.924 to 0.961. The 95% limits of agreement were −19.90 to 35.56 µm between Sirius and RTVue, −10.69 to 35.85 µm between Sirius and AL-Scan, and −4.07 to 13.58 µm between RTVue and AL-Scan. Conclusion: Different results could be obtained through different noncontact devices in CCT measurements. Although the measurement values obtained by these devices show a high level of correlation, it would be a more correct approach to not use them directly interchangeably in clinical practice. Evaluation and follow-up of CCT should be performed using the same device.


2012 ◽  
Vol 2012 ◽  
pp. 1-5 ◽  
Author(s):  
Ilona Piotrowiak ◽  
Beata Soldanska ◽  
Mateusz Burduk ◽  
Bartlomiej J. Kaluzny ◽  
Jozef Kaluzny

Background and Objective. Evaluation of agreement, repeatability, and reproducibility of central and minimal corneal thickness (CCT and MCT) measurements obtained by SOCT, the Scheimpflug system, and ultrasound pachymetry. Materials and Methods. 28 eyes of healthy patients were enrolled. Pachymetry measurements were performed with SOCT, the Scheimpflug system, and ultrasound instrument. Each measurement was taken by 3 operators on 3 devices providing a total of 2100 measurements. Results. The mean CCT for SOCT, Scheimpflug system, and ultrasound instrument was 537.92, 545.94, and 555.74 μm, respectively, (). The respective mean coefficients of repeatability for CCT were 0.61, 0.82 and 0.80, whereas mean coefficients of interoperator reproducibility for CCT were 0.91, 1.11, and 1.25. Conclusions. CCT and MCT measurements show moderate agreement between instruments. The repeatability and interoperator reproducibility of the results obtained by SOCT are somewhat higher. The operator's impact on CCT and MCT measurements is insignificant in all devices.


2019 ◽  
Author(s):  
Anqi Liu ◽  
Zequan Xu ◽  
Jingjing Wang ◽  
Feng Liu ◽  
Yanming Jiang ◽  
...  

Abstract Purpose To assess the precision (repeatability and reproducibility) and agreement of central corneal thickness (CCT) measurements by E-pach, Pentacam, Corvis ST and A-scan devices in healthy eyes. Methods Instrument diagnostic test study. A total of 100 patients were enrolled to measure right-eye CCT by E-pach, Pentacam, Corvis ST, and A-scan devices. To assess repeatability and reproducibility, the test-retest repeatability (TRT) and intraclass correlation coefficient (ICC) were calculated. The agreement among the four devices was evaluated with Bland-Altman plots. Results The E-pach showed repeatability (ICC=0.9981), interobserver reproducibility (ICC=0.9971) and intersession reproducibility (ICC=0.9825); the Pentacam, Corvis ST and A-scan also showed similar repeatability (all ICC≥0.9679), interobserver reproducibility (all ICC≥0.9730) and intersession reproducibility (all ICC≥0.9647). However, the E-pach yielded CCT values that were obviously lower than those of the A-scan and Pentacam (P<0.001) but higher than those of the Corvis ST (P<0.001). The 95% limits of agreement (LoA) in the Bland-Altman plots were 44.5 µm (E-pach vs. A-scan), 34.9 µm (E-pach vs. Corvis CT) and 32.5 µm (E-pach vs. Pentacam). Conclusions The E-pach is a portable, reliable and inexpensive pachymeter. However, the CCT values obtained from the E-pach are not interchangeable with those from Pentacam, Corvis ST and A-scan devices.


2019 ◽  
Vol 35 (3) ◽  
Author(s):  
Munira Shakir, Ronak Afza Memon Sahira Wasim Shakir Zafar

Purpose: To determine the mean difference of central corneal thickness measurements by using ultrasound Pachymetry and oculus Wave light Occulyzer II. Study Design: Cross-sectional comparative study using non-probability consecutive sampling. Place and Duration of Study: This study was conducted at Department of Ophthalmology, Liaquat National Hospital Karachi from November 2018 to June 2019. Material & Methods: After approval from ethical committee, patients were included in our study on the basis of inclusion & exclusion criteria. Central corneal thickness measurements were taken by using ultrasound Pachymetry & oculus Wave light Occulyzer II. All the data was collected by single researcher. The results were plotted, compared & analyzed. Paired t-test was used for the comparison of quantitative variables. Results: There were 130 patients included in the study out of which 73 (56.2%) were males and 57 (43.8%) were females. Mean age of these patients was 33.9 � 8.9 years. The mean � SD thinnest oculus Wave light Occulyzer II measurement was 538.61 � 23.46 ?m and ranged between 476.0 and 619.0 ?m whereas the mean thinnest ultrasound Pachymetry measurement was 535.1 � 21.816 ?m and ranged between 482 and 601 ?m. There was highly significant correlation of central corneal thickness between both the instruments. (r = 0.96, p < 0.001). Conclusion: There is a high correlation of central corneal thickness between the readings obtained from ultrasound and optical pachymetry machines therefore oculus Wave light Occulyzer II can be used as an alternative technique to ultrasound pachymetry while assessing CCT in clinical settings. Key Words: Central corneal thickness, Ultrasound Pachymetry, Optical Pachymetry.


Sign in / Sign up

Export Citation Format

Share Document