scholarly journals Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure

2013 ◽  
Vol 2013 ◽  
pp. 1-5 ◽  
Author(s):  
Fusako Fujimura ◽  
Kazutaka Kamiya ◽  
Kazuko Fujiwara ◽  
Nobuyuki Shoji ◽  
Kimiya Shimizu

Purpose. To investigate the repeatability and reproducibility of intraocular pressure (IOP) and central corneal thickness (CCT) measurements using a noncontact tono/pachymeter (NT-530P) and to assess the correlation of CCT with IOP.Methods. Forty-six eyes of healthy volunteers were measured by two examiners. Three consecutive measurements per eye were performed. Repeatability was assessed using the coefficient of variation, and reproducibility was assessed using Bland-Altman plots. Linear correlations were used to determine agreement between CCT and noncorrected IOP and CCT and corrected IOP, which was calculated using a formula built into the NT-530P.Results. The coefficient of variation for IOP was 6.4% and for CCT was 0.4%. The 95% limits of agreement between examiners were−0.17±1.42 mmHg (range:−2.95to 2.61 mmHg) for IOP,−0.93±4.37 μm (range:−9.50to 7.64 μm) for CCT. The corrected IOP was significantly higher than the noncorrected IOP (P=0.010.3). The noncorrected IOP significantly correlated with CCT (r=−0.4883,P=0.0006). The corrected IOP showed no significant correlation with CCT (r=−0.0285,P=0.8509).Conclusions. NT-530P offered repeatability and reproducibility in both IOP and CCT measurements. The corrected IOP calculated using the NT-530P was independent of the CCT, suggesting that this IOP may be less influenced by the central corneal thickness.

2015 ◽  
Vol 2015 ◽  
pp. 1-8 ◽  
Author(s):  
Ayong Yu ◽  
Weiqi Zhao ◽  
Giacomo Savini ◽  
Zixu Huang ◽  
Fangjun Bao ◽  
...  

Purpose.To assess the repeatability and reproducibility of central corneal thickness (CCT) measurements by corneal dynamic Scheimpflug analyzer Corvis ST in normal eyes and compare the agreement with Pentacam rotating Scheimpflug System and ultrasound pachymetry.Methods.84 right eyes underwent Corvis ST measurements performed by two operators. The test-retest repeatability (TRT), within-subject coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were used to evaluate the intraoperator repeatability and interoperator reproducibility. CCT measurements also were obtained from Pentacam and ultrasound pachymetry by the first operator. The agreement between the three devices was evaluated with 95% limits of agreement (LoA) and Bland-Altman plots.Results.Corvis ST showed high repeatability as indicated by TRT ≤ 13.0 μm, CoV < 0.9%, and ICC > 0.97. The interoperator reproducibility was also excellent. The CoV was <0.9%, and ICC was >0.97. Corvis ST showed significantly lower values than Pentacam and ultrasound pachymetry (P<0.001). The 95% LoA between Corvis ST and Pentacam or ultrasound pachymetry were −15.8 to 9.5 μm and −27.9 to 12.3 μm, respectively.Conclusions.Corvis ST showed excellent repeatability and interoperator reproducibility of CCT measurements in normal eyes. Corvis ST is interchangeable with Pentacam but not with ultrasound pachymetry.


2019 ◽  
Vol 30 (2) ◽  
pp. 258-263 ◽  
Author(s):  
Barbara Cvenkel ◽  
Makedonka Atanasovska Velkovska ◽  
Vesna Dimovska Jordanova

Purpose: To evaluate and compare the accuracy of self-measurement of intraocular pressure using Icare Home rebound tonometer with Goldmann applanation tonometer and assess acceptability of self-tonometry in patients with glaucoma and ocular hypertension. Methods: In the study, 117 subjects were trained to use Icare Home for self-measurement. Icare Home tonometer readings were compared with Goldmann applanation tonometer, including one eye per patient. Agreement between the two methods of measurement was evaluated by Bland and Altmann analysis. Questionnaire was used to evaluate patients’ perception of self-tonometry. Results: One hundred and three out of 117 patients (88%) were able to measure their own intraocular pressure and 96 (82%) fulfilled the requirements for certification. The mean (SD) difference Goldmann applanation tonometer minus Icare Home was 1.2 (2.4) mmHg (95% limits of agreement, –3.4 to 5.9 mmHg). The magnitude of bias between the two methods depended on central corneal thickness, with greater bias at central corneal thickness <500 µm. In 65 out of 96 subjects (67.7%), Icare Home results were within 2 mmHg of the Goldmann applanation tonometer. Seventy-three out of 93 (78.5%) felt that self-tonometry was easy to use and 75 patients (80.6%) responded that they would use the device at home. Conclusion: Icare Home tonometry tends to slightly underestimate intraocular pressure compared to Goldmann applanation tonometer. Most patients were able to perform self-tonometry and found it acceptable for home use. Measurements using rebound self-tonometry could improve the quality of intraocular pressure data and optimize treatment regimen.


2020 ◽  
Vol 44 (2) ◽  
pp. 98-103
Author(s):  
Mehtap Sagus Aydin ◽  
◽  
Ayse Ipek Akyuz Unsal ◽  
Sayime Aydin Eroglu ◽  
Mustafa Unubol ◽  
...  

2017 ◽  
Vol 2017 ◽  
pp. 1-4 ◽  
Author(s):  
Bernardo T. Lopes ◽  
Cynthia J. Roberts ◽  
Ahmed Elsheikh ◽  
Riccardo Vinciguerra ◽  
Paolo Vinciguerra ◽  
...  

Purpose. To assess the repeatability and reproducibility of dynamic corneal response parameters measured by the Corvis ST (Oculus, Wetzlar, Germany). Methods. One eye randomly selected from 32 healthy volunteers was examined by the Corvis ST. Three different devices were used in an alternated random order for taking three measurements at each device in each subject. Standard intraocular pressure (IOP), the biomechanical-compensated IOP (bIOP), and DCR parameters were evaluated. The within-subject standard deviation (ζw) and coefficient of variation (CV) were assessed. Results. Regarding pressure indices, the ζw was below 1 mmHg for repeatability (0.98 for IOP and 0.89 for bIOP) and the CV was 6.6% for IOP and 6.1% for bIOP. For reproducibility, the ζw was around 1 mmHg (1.12 for IOP and 1.05 for bIOP) and the CV was 7.6% for IOP and 7.1% for bIOP. Most of DCR indices presented CV for repeatability below 4%. For reproducibility, the CV of most of the indices were below 6%. The deformation amplitude (DA) ratio in 1 mm and integrated radius were below 4% (1.2% and 3.8%, resp.). Conclusions. The Corvis ST showed good precision (repeatability and reproducibility) for IOP measurements and for DCR in healthy eyes.


2019 ◽  
Author(s):  
Anqi Liu ◽  
Zequan Xu ◽  
Jingjing Wang ◽  
Feng Liu ◽  
Yanming Jiang ◽  
...  

Abstract Purpose To assess the precision (repeatability and reproducibility) and agreement of central corneal thickness (CCT) measurements by E-pach, Pentacam, Corvis ST and A-scan devices in healthy eyes. Methods Instrument diagnostic test study. A total of 100 patients were enrolled to measure right-eye CCT by E-pach, Pentacam, Corvis ST, and A-scan devices. To assess repeatability and reproducibility, the test-retest repeatability (TRT) and intraclass correlation coefficient (ICC) were calculated. The agreement among the four devices was evaluated with Bland-Altman plots. Results The E-pach showed repeatability (ICC=0.9981), interobserver reproducibility (ICC=0.9971) and intersession reproducibility (ICC=0.9825); the Pentacam, Corvis ST and A-scan also showed similar repeatability (all ICC≥0.9679), interobserver reproducibility (all ICC≥0.9730) and intersession reproducibility (all ICC≥0.9647). However, the E-pach yielded CCT values that were obviously lower than those of the A-scan and Pentacam (P<0.001) but higher than those of the Corvis ST (P<0.001). The 95% limits of agreement (LoA) in the Bland-Altman plots were 44.5 µm (E-pach vs. A-scan), 34.9 µm (E-pach vs. Corvis CT) and 32.5 µm (E-pach vs. Pentacam). Conclusions The E-pach is a portable, reliable and inexpensive pachymeter. However, the CCT values obtained from the E-pach are not interchangeable with those from Pentacam, Corvis ST and A-scan devices.


2021 ◽  
Vol 8 (33) ◽  
pp. 3084-3089
Author(s):  
Ram Kumar Jaiswal ◽  
Ramyash Singh Yadav ◽  
Mridula Ranjan ◽  
Dipti Wahi ◽  
Chiranji Rai

BACKGROUND Myopia is a complex trait including both genetic and environmental factors as well as gene-environment interactions. It has been recognized as a risk factor for development of glaucoma. Both Myopia and raised IOP are known to increase mechanical stress at optic nerve head leading to glaucomatous nerve damage. This study intends to evaluate the clinical profile of myopic subjects and its correlation with Primary open angle glaucoma (POAG). METHODS A prospective observational study done from November 2019 to December 2020 after taking proper informed consent and ethical clearance. 164 eyes of 96 patients studied were divided into three groups, Group 1: low myopia (< -3.00D), Group 2: moderate myopia (-3.00 D to -5.75 D), Group 3: high myopia (≥ -6.00 D). A complete ocular examination was performed. Intraocular pressure was measured using Goldmann applanation tonometer. Visual field analysis using Humphrey automated perimetry was done in patients with suspected primary open angle glaucoma (POAG). Angle parameters and central corneal thickness (CCT) were measured using anterior segment optical coherence tomography (AS-OCT). RESULTS 164 eyes of 96 Myopic subjects were studied with no dropout during study period. Mean age was 46.05 yr. (range: 25-75 yr.). The refraction ranged from -0.50 DS to -17.00 DS. There was no statistically significant difference between Intraocular pressure (IOP), Central corneal thickness (CCT), corrected IOP and Nasal and Temporal Trabecular-iris Angle (TIA) between male and female of same age group. Mean IOP and mean CCT were found to vary significantly with age and with higher degree of myopia. Corrected IOP, Nasal and Temporal TIA increase significantly with higher degree of myopia. Cup-disc ratio (CDR) was found to be significantly higher in patients with moderate to high degree of Myopia. CONCLUSIONS Myopia is an important risk factor for development of primary open angle glaucoma, with its incidence increasing in patients with moderate to high myopia. KEYWORDS Myopia, Primary Open Angle Glaucoma, Intraocular Pressure, Central Corneal Thickness, Trabecular Iris Angle


Clinics ◽  
2010 ◽  
Vol 65 (10) ◽  
pp. 967-970 ◽  
Author(s):  
Rafael Lacerda Furlanetto ◽  
Antonio Carlos Facio Jr ◽  
Marcelo Hatanaka ◽  
Remo Susanna Junior

1970 ◽  
Vol 12 (3) ◽  
pp. 156-158
Author(s):  
Ozlem Gurses Sahin

Aim: To evaluate whether a significant difference exists between the initial and final mean central corneal thickness that might affect Goldmann applanation tonometer recordings of patients with allergic conjunctivitis treated with desonide 0.25% ophthalmic solution, and to determine the effect of desonide on intraocular pressure by using adjusted intraocular pressure values.Methods: This double-blind randomised placebo-controlled trial enrolled 30 patients (60 eyes) with allergic conjunctivitis. Patients were randomly assigned to receive desonide to 1 eye (study eye) and preservative-free tear supplement (Tears Naturale Free®) to the other eye (control eye) 3 times daily for 3 weeks. Goldmann applanation tonometry and central corneal thickness of both eyes were recorded on the day of presentation and 3 weeks later.Results: Regression of itching, tearing, conjunctival hyperaemia, and tarsal conjunctival papillary reaction were noted in the study eyes after 3 weeks of treatment. The control eyes showed mild regression of pruritis, tearing, conjunctival hyperaemia, and tarsal conjunctival papillary reaction after 3 weeks. There was a significant difference between the initial and final mean central corneal thickness values for the study eyes (p = 0.003). There were no statistically significant differences between the initial and final mean values of Goldmann applanation tonometry and adjusted intraocular pressure for the study eyes. No statistically significant differences were found between the initial and final mean values of any of the parameters for the control eyes.Conclusion: Inflammation-induced increase in central corneal thickness of patients with allergic conjunctivitis treated with desonide showed statistically significant regression. However, this regression did not significantly affect Goldmann applanation tonometry and adjusted intraocular pressure values of the treated eyes.


Sign in / Sign up

Export Citation Format

Share Document