SU-E-J-137: A Theoretical Model for the Analysis of Residual Motion Artifacts in 4D CT Scans

2012 ◽  
Vol 39 (6Part8) ◽  
pp. 3684-3684
Author(s):  
T Kim ◽  
J Yoon ◽  
S Kang ◽  
T Suh
2021 ◽  
Vol 42 (6) ◽  
pp. 1805-1828
Author(s):  
Daniele Mascali ◽  
Marta Moraschi ◽  
Mauro DiNuzzo ◽  
Silvia Tommasin ◽  
Michela Fratini ◽  
...  

2021 ◽  
Vol 11 (1) ◽  
Author(s):  
Doil Kim ◽  
Jiyoung Choi ◽  
Duhgoon Lee ◽  
Hyesun Kim ◽  
Jiyoung Jung ◽  
...  

AbstractA novel motion correction algorithm for X-ray lung CT imaging has been developed recently. It was designed to perform for routine chest or thorax CT scans without gating, namely axial or helical scans with pitch around 1.0. The algorithm makes use of two conjugate partial angle reconstruction images for motion estimation via non-rigid registration which is followed by a motion compensated reconstruction. Differently from other conventional approaches, no segmentation is adopted in motion estimation. This makes motion estimation of various fine lung structures possible. The aim of this study is to explore the performance of the proposed method in correcting the lung motion artifacts which arise even under routine CT scans with breath-hold. The artifacts are known to mimic various lung diseases, so it is of great interest to address the problem. For that purpose, a moving phantom experiment and clinical study (seven cases) were conducted. We selected the entropy and positivity as figure of merits to compare the reconstructed images before and after the motion correction. Results of both phantom and clinical studies showed a statistically significant improvement by the proposed method, namely up to 53.6% (p < 0.05) and up to 35.5% (p < 0.05) improvement by means of the positivity measure, respectively. Images of the proposed method show significantly reduced motion artifacts of various lung structures such as lung parenchyma, pulmonary vessels, and airways which are prominent in FBP images. Results of two exemplary cases also showed great potential of the proposed method in correcting motion artifacts of the aorta which is known to mimic aortic dissection. Compared to other approaches, the proposed method provides an excellent performance and a fully automatic workflow. In addition, it has a great potential to handle motions in wide range of organs such as lung structures and the aorta. We expect that this would pave a way toward innovations in chest and thorax CT imaging.


2016 ◽  
Vol 119 (1) ◽  
pp. 172-178 ◽  
Author(s):  
Jin-Zhi Wang ◽  
Jian-Bin Li ◽  
Huan-Peng Qi ◽  
Yan-Kang Li ◽  
Yue Wang ◽  
...  

Author(s):  
B.J. Slotman ◽  
F.J. Lagerwaard ◽  
R.W. Underberg ◽  
J.P. Cuijpers ◽  
J.R. van Sornsen de Koste ◽  
...  

Author(s):  
G.F. Persson ◽  
D.E. Nygaard ◽  
P. Munck af Rosenschöld ◽  
S. Korreman ◽  
L. Specht

2013 ◽  
Vol 40 (6Part26) ◽  
pp. 434-435
Author(s):  
E Aliotta ◽  
D Thomas ◽  
S Gaudio ◽  
B White ◽  
S Jani ◽  
...  

Sign in / Sign up

Export Citation Format

Share Document