Magnetic Resonance Imaging Based Flow Field and Lagrangian Particle Tracking From a Left Ventricular Assist Device

2019 ◽  
Vol 142 (2) ◽  
Author(s):  
Omid Amili ◽  
Robroy MacIver ◽  
Filippo Coletti

Abstract This study explores the optimal left ventricular assist device (LVAD) cannula outflow configuration in a patient-specific replica of the aorta. The volumetric velocity field is measured using phase-contrast magnetic resonance imaging (PC-MRI) under a physiologically relevant steady flow. The effect of the LVAD outflow graft insertion site and anastomosis angle on the transport of embolic particles to cranial vessels is studied by solving the particle equation of motion for spheres in the range of 0.1–1.0 mm using the measured three-dimensional (3D) velocity field. Results show that for a given aorta anatomy, it is possible to design the cannula graft location and terminal curvature so that the probability of embolic transport to the cranial vessels is significantly minimized. This is particularly important since the complex flow pattern in each cannula case affects the embolic trajectories differently, and hence the common assumption that particles distribute by the volumetric flow division does not hold.

Author(s):  
J P Cassella ◽  
V Salih ◽  
T R Graham

Left ventricular assist systems are being developed for eventual long term or permanent implantation as an alternative to heart transplantation in patients unsuitable for or denied the transplant option. Evaluation of the effects of these devices upon normal physiology is required. A preliminary study was conducted to evaluate the morphology of aortic tissue from calves implanted with a pneumatic Left Ventricular Assist device-LVAD. Two 3 month old heifer calves (calf 1 and calf 2) were electively explanted after 128 days and 47 days respectively. Descending thoracic aortic tissue from both animals was removed immediately post mortem and placed into karnovsky’s fixative. The tissue was subsequently processed for transmission electron microscopy (TEM). Some aortic tissue was fixed in neutral buffered formalin and processed for routine light microscopy.


Sign in / Sign up

Export Citation Format

Share Document