Pulse-Decay Method for Measuring the Thermal Conductivity of Living Tissues

1981 ◽  
Vol 103 (4) ◽  
pp. 253-260 ◽  
Author(s):  
M. M. Chen ◽  
K. R. Holmes ◽  
V. Rupinskas

The present communication presents a single microprobe technique for measuring tissue thermal properties based on the dissipation of a measured amount of energy and the observation of the resulting temperature rise a given time later. An advantage of this method is that the effective sampling volume can be varied by varying the measurement time. Using a measurement time of a few seconds, the sampling volume was estimated to be several orders of magnitude greater than the probe volume. Hence artifacts due to probe-induced trauma or stress would be insignificant. Additional advantages of the technique are: the results were independent of the probe shape, size and properties, and hence represents absolute measurements without the need for calibration; the required electronics and computations are simple; the determination of thermal conductivity requires only a single measurement; and comparison of data at different measurement times yields a clear and unequivocal indication of nonconductive contributions of heat transfer, if present.

Author(s):  
Marcelo Borges dos Santos ◽  
CLAUDIA BITTENCOURT ◽  
Ana Carolina Mendonça Mansur ◽  
Luís Mauro Moura ◽  
Carlos Augusto Castro Ferreira

Sign in / Sign up

Export Citation Format

Share Document