Ultrastructural Localization of Ions. Iii. Distribution of Chloride in Mesophyll Cells of Mangrove (Aegiceras Corniculatum Blanco).

1976 ◽  
Vol 3 (3) ◽  
pp. 367 ◽  
Author(s):  
RFMV Steveninck ◽  
WD Armstrong ◽  
PD Peters ◽  
TA Hall

Ultrastructural localization of Cl- by means of a silver ion precipitation technique combined with X-ray energy spectroscopy (EMMA-4 electron microscope plus KEVEX energy dispersive analyser) showed the presence of two types of vacuoles in mesophyll cells of mangrove (A. corniculatum Blanco). Type A vacuoles contained large amounts of osmiophilic organic solute and little or no Cl-, while type B vacuoles were shown to be free of osmiophilic organic solute but to contain significant quantities of Cl-. X-ray spectroscopy of electron-opaque deposits in chloroplasts showed the presence of AgCl deposits and deposits of silver in approximately equal proportions. Fine granular deposits in plastoglobuli were also shown to consist of silver. Deposits in the cytoplasmic phase and in the plasmodesmata, however, always consisted of AgCl. The significance of the two types of vacuoles is discussed in relation to the symplastic or apoplastic transport of Cl- from conducting tissue to the salt glands.

1976 ◽  
Vol 3 (3) ◽  
pp. 359 ◽  
Author(s):  
RFMV Steveninck ◽  
B Ballment ◽  
PD Peters ◽  
TA Hall

The ultrastructural localization of Cl- by means of a silver precipitation technique has been verified by means of X-ray analytical electron microscopy (EMMA-4 electron microscope plus KEVEX energy dispersive analyser). It was shown that the localization method was specific for Cl- in the case of electron-dense deposits in chloroplasts, plasmodesmata and plasmalemma of barley mesophyll cells. It was pointed out that the use of nitric acid rinses in order to remove non-specific deposits of silver phosphates and carbonates has pitfalls and that the precipitation technique should always be used in conjunction with X-ray microprobe analysis.


1990 ◽  
Vol 68 (3) ◽  
pp. 646-650 ◽  
Author(s):  
I. Ockenden ◽  
J. N. A. Lott

Embryos of Cucurbita maxima, Cucurbita andreana, and their reciprocal hybrids store phosphorus, magnesium, potassium, and calcium in distinctive globoid crystals within protein bodies found in cotyledon mesophyll cells. Samples of cotyledon tissue from each of the four kinds of embryos were cryogenically prepared and analyzed by X-ray microanalysis in the frozen state in a scanning electron microscope. The relative proportions of phosphorus, magnesium, potassium, and calcium were expressed as peak to background ratios. Total concentrations of the four elements in whole cotyledons, as measured by neutron activation analysis, were similar to the trends shown by the peak to background ratios for the same elements. This supports the concept that the bulk of the elements P, K, Mg, and Ca in seed tissues is stored in globoid crystals.


Author(s):  
G. Cliff ◽  
M.J. Nasir ◽  
G.W. Lorimer ◽  
N. Ridley

In a specimen which is transmission thin to 100 kV electrons - a sample in which X-ray absorption is so insignificant that it can be neglected and where fluorescence effects can generally be ignored (1,2) - a ratio of characteristic X-ray intensities, I1/I2 can be converted into a weight fraction ratio, C1/C2, using the equationwhere k12 is, at a given voltage, a constant independent of composition or thickness, k12 values can be determined experimentally from thin standards (3) or calculated (4,6). Both experimental and calculated k12 values have been obtained for K(11<Z>19),kα(Z>19) and some Lα radiation (3,6) at 100 kV. The object of the present series of experiments was to experimentally determine k12 values at voltages between 200 and 1000 kV and to compare these with calculated values.The experiments were carried out on an AEI-EM7 HVEM fitted with an energy dispersive X-ray detector.


Author(s):  
R. Sinclair ◽  
B.E. Jacobson

INTRODUCTIONThe prospect of performing chemical analysis of thin specimens at any desired level of resolution is particularly appealing to the materials scientist. Commercial TEM-based systems are now available which virtually provide this capability. The purpose of this contribution is to illustrate its application to problems which would have been intractable until recently, pointing out some current limitations.X-RAY ANALYSISIn an attempt to fabricate superconducting materials with high critical currents and temperature, thin Nb3Sn films have been prepared by electron beam vapor deposition [1]. Fine-grain size material is desirable which may be achieved by codeposition with small amounts of Al2O3 . Figure 1 shows the STEM microstructure, with large (∽ 200 Å dia) voids present at the grain boundaries. Higher quality TEM micrographs (e.g. fig. 2) reveal the presence of small voids within the grains which are absent in pure Nb3Sn prepared under identical conditions. The X-ray spectrum from large (∽ lμ dia) or small (∽100 Ǻ dia) areas within the grains indicates only small amounts of A1 (fig.3).


Author(s):  
M.D. Ball ◽  
H. Lagace ◽  
M.C. Thornton

The backscattered electron coefficient η for transmission electron microscope specimens depends on both the atomic number Z and the thickness t. Hence for specimens of known atomic number, the thickness can be determined from backscattered electron coefficient measurements. This work describes a simple and convenient method of estimating the thickness and the corrected composition of areas of uncertain atomic number by combining x-ray microanalysis and backscattered electron intensity measurements.The method is best described in terms of the flow chart shown In Figure 1. Having selected a feature of interest, x-ray microanalysis data is recorded and used to estimate the composition. At this stage thickness corrections for absorption and fluorescence are not performed.


Author(s):  
W. Brünger

Reconstructive tomography is a new technique in diagnostic radiology for imaging cross-sectional planes of the human body /1/. A collimated beam of X-rays is scanned through a thin slice of the body and the transmitted intensity is recorded by a detector giving a linear shadow graph or projection (see fig. 1). Many of these projections at different angles are used to reconstruct the body-layer, usually with the aid of a computer. The picture element size of present tomographic scanners is approximately 1.1 mm2.Micro tomography can be realized using the very fine X-ray source generated by the focused electron beam of a scanning electron microscope (see fig. 2). The translation of the X-ray source is done by a line scan of the electron beam on a polished target surface /2/. Projections at different angles are produced by rotating the object.During the registration of a single scan the electron beam is deflected in one direction only, while both deflections are operating in the display tube.


Author(s):  
Zenji Horita ◽  
Ryuzo Nishimachi ◽  
Takeshi Sano ◽  
Minoru Nemoto

Absorption correction is often required in quantitative x-ray microanalysis of thin specimens using the analytical electron microscope. For such correction, it is convenient to use the extrapolation method[l] because the thickness, density and mass absorption coefficient are not necessary in the method. The characteristic x-ray intensities measured for the analysis are only requirement for the absorption correction. However, to achieve extrapolation, it is imperative to obtain data points more than two at different thicknesses in the identical composition. Thus, the method encounters difficulty in analyzing a region equivalent to beam size or the specimen with uniform thickness. The purpose of this study is to modify the method so that extrapolation becomes feasible in such limited conditions. Applicability of the new form is examined by using a standard sample and then it is applied to quantification of phases in a Ni-Al-W ternary alloy.The earlier equation for the extrapolation method was formulated based on the facts that the magnitude of x-ray absorption increases with increasing thickness and that the intensity of a characteristic x-ray exhibiting negligible absorption in the specimen is used as a measure of thickness.


Author(s):  
S. M. Zemyan ◽  
D. B. Williams

As has been reported elsewhere, a thin evaporated Cr film can be used to monitor the x-ray peak to background ratio (P/B) in an analytical electron microscope. Presented here are the results of P/B measurements for the Cr Ka line on a Philips EM430 TEM/STEM, with Link Si(Li) and intrinsic Ge (IG) x-ray detectors. The goal of the study was to determine the best conditions for x-ray microanalysis.We used the Fiori P/B definition, in which P/B is the ratio of the total peak integral to the average background in a 10 eV channel beneath the peak. Peak and background integrals were determined by the window method, using a peak window from 5.0 to 5.7 keV about Cr Kα, and background windows from 4.1 to 4.8 keV and 6.3 to 7.0 keV.


Author(s):  
Marc H. Peeters ◽  
Max T. Otten

Over the past decades, the combination of energy-dispersive analysis of X-rays and scanning electron microscopy has proved to be a powerful tool for fast and reliable elemental characterization of a large variety of specimens. The technique has evolved rapidly from a purely qualitative characterization method to a reliable quantitative way of analysis. In the last 5 years, an increasing need for automation is observed, whereby energy-dispersive analysers control the beam and stage movement of the scanning electron microscope in order to collect digital X-ray images and perform unattended point analysis over multiple locations.The Philips High-speed Analysis of X-rays system (PHAX-Scan) makes use of the high performance dual-processor structure of the EDAX PV9900 analyser and the databus structure of the Philips series 500 scanning electron microscope to provide a highly automated, user-friendly and extremely fast microanalysis system. The software that runs on the hardware described above was specifically designed to provide the ultimate attainable speed on the system.


Author(s):  
Jean-Claude Jésior ◽  
Roger Vuong ◽  
Henri Chanzy

Starch is arranged in a crystalline manner within its storage granules and should thus give sharp X-ray diagrams. Unfortunately most of the common starch granules have sizes between 1 and 100μm, making them too small for an X-ray study on individual grains. There is only one instance where an oriented X-ray diagram could be obtained on one sector of an individual giant starch granule. Despite their small size, starch granules are still too thick to be studied by electron diffraction with a transmission electron microscope. The only reported study on starch ultrastructure using electron diffraction on frozen hydrated material was made on small fragments. The present study has been realized on thin sectioned granules previously litnerized to improve the signal to noise ratio.Potato starch was hydrolyzed for 10 days in 2.2N HCl at 35°C, dialyzed against water until neutrality and embedded in Nanoplast. Sectioning was achieved with a commercially available low-angle “35°” diamond knife (Diatome) after a very carefull trimming and a pre-sectioning with a classical “45°” diamond knife. Sections obtained at a final sectioning angle of 42.2° (compared with the usual 55-60°) and at a nominal thickness of 900Å were collected on a Formvar-carbon coated grid. The exact location of the starch granules in their sections was recorded by optical microscopy on a Zeiss Universal polarizing microscope (Fig. 1a). After rehydration at a relative humidity of 95% for 24 hours they were mounted on a Philips cryoholder and quench frozen in liquid nitrogen before being inserted under frozen conditions in a Philips EM 400T electron microscope equipped with a Gatan anticontaminator and a Lhesa image intensifier.


Sign in / Sign up

Export Citation Format

Share Document