Oxygen-17 nuclear quadrupole double resonance spectroscopy

1978 ◽  
Vol 13 ◽  
pp. 75 ◽  
Author(s):  
Theodore L. Brown ◽  
C. P. Cheng
1990 ◽  
Vol 45 (3-4) ◽  
pp. 595-598 ◽  
Author(s):  
M. M. P. Khurshid ◽  
F. Zhenye ◽  
J.A.S. Smith

Abstract This paper discusses the design of a variable-temperature high-sensitivity field cycling spectrometer for nuclear magnetic resonance and nuclear quadrupole double resonance experiments. The instrument is capable of detecting N-14 quadrupole resonance signals from samples as small as 0.1 g down to temperatures of 150 K. The magnetic field cycling is performed by mechanical transfer of the sample using a pair of switchable electromagnetic solenoids, in times of about 0.2 s.