scholarly journals A Comparison Study Between Point-of-Care Testing Systems and Central Laboratory for Determining Blood Glucose in Venous Blood

2016 ◽  
Vol 31 (3) ◽  
pp. e22051 ◽  
Author(s):  
Huiping Wei ◽  
Fang Lan ◽  
Qitian He ◽  
Haiwei Li ◽  
Fuyong Zhang ◽  
...  
2020 ◽  
Vol 58 (3) ◽  
pp. 445-455
Author(s):  
Annette Baumstark ◽  
Nina Jendrike ◽  
Ulrike Kamecke ◽  
Christina Liebing ◽  
Stefan Pleus ◽  
...  

AbstractBackgroundThe professional-use systems HemoCue® Glucose 201+ (HC201+) and HemoCue® Glucose 201 RT (HC201RT) are widely used for point-of-care testing (POCT) of blood glucose (BG). HC201RT utilizes unit-use microcuvettes which can be stored at room temperature, whereas HC201+ microcuvettes have to be stored at <8 °C. In this study, system accuracy of HC201+ and HC201RT was evaluated using capillary and venous blood samples.MethodsFor each system, two reagent system lots were evaluated within a period of 2 years based on testing procedures of ISO 15197:2013, a standard applicable for self-monitoring of blood glucose (SMBG) systems. For each reagent system lot, the investigation was performed by using 100 capillary and 95 to 99 venous blood samples. Comparison measurements were performed with a hexokinase laboratory method. Accuracy criteria of ISO 15197:2013 and POCT12-A3 were applied. In addition, bias was analyzed according to Bland and Altman, and error grid analysis was performed.ResultsWhen measuring capillary samples, both systems fulfilled accuracy requirements of ISO 15197:2013 and POCT12-A3 with the investigated reagent system lots. When measuring venous samples, only HC201+ fulfilled these requirements. Bias between HC201+ and reference measurements was more consistent over venous and capillary samples and microcuvette lots than for HC201RT. Error grid analysis showed that clinical actions might have been different depending on which system was used.ConclusionsIn this study, HC201+ showed a high level of accuracy irrespective of the sample type (capillary or venous). In contrast, HC201RT measurement results were markedly affected by the type of sample.


Author(s):  
Ellis Jacobs ◽  
Jerzy W. Naskalski ◽  
Niels Fogh-Andersen ◽  
Christoph Ritter ◽  
Andrezj Lewenstam ◽  
...  

Author(s):  
Julie D Newman ◽  
Nerissa S Pecache ◽  
Charles P Barfield ◽  
Nicholas DH Balazs

Background: Blood glucose measurements in newborns at risk of hypoglycaemia are an essential part of their medical management. Blood glucose measurements obtained by point-of-care testing using an AVL Omni 9 blood gas and metabolite analyser were compared with those obtained in the central laboratory using a Dade Dimension RXL analyser. Methods: Blood glucose was measured at the point of care by nursing staff using an AVL Omni 9 blood gas and metabolite analyser and compared to results obtained in the central laboratory using a DADE Dimension RXL analyser. In total, 123 samples were taken from 114 babies admitted to the neonatal unit. Results: The limits of agreement between the AVL Omni 9 and the central laboratory were 0·0±0·6 mmol/L for glucose values between 0·5 and 13 mmol/L. Regression analysis showed: AVL Omni 9 glucose = 0·977 x plasma glucose+0·14. There was also a good correlation ( r=0·92) between the AVL Omni 9 and the DADE Dimension RXL analyser for glucose values <3 mmol/L. The limits of agreement for the AVL Omni 9 when compared with the DADE Dimension RXL analyser were -0·1±0·5 mmol/L. Discussion: Point-of-care testing of blood glucose using the AVL Omni 9 blood gas and metabolite analyser is a reliable means of measuring blood glucose and has the advantage of providing a fast result using small volumes of blood.


Author(s):  
Gomathi Ramaswamy ◽  
Kashish Vohra ◽  
Kapil Yadav ◽  
Ravneet Kaur ◽  
Tripti Rai ◽  
...  

Abstract Introduction Globally around 47.4% of children and in India, 58% of children aged 6–59 months are anemic. Diagnosis of anemia in children using accurate technologies and providing adequate treatment is essential to reduce the burden of anemia. Point-of-care testing (POCT) devices is a potential option for estimation of hemoglobin in peripheral and field settings were the hematology analyzer and laboratory services are not available. Objectives To access the validity of the POCTs (invasive and non-invasive devices) for estimation of hemoglobin among children aged 6–59 months compared with hematology analyzer. Methods The study participants were enrolled from the pediatric outpatient department in Haryana, India, from November 2019 to January 2020. Hemoglobin levels of the study participants were estimated in Sahli’s hemoglobinometer and invasive digital hemoglobinometers (DHs) using capillary blood samples. Hemoglobin levels in non-invasive DH were assessed from the finger/toe of the children. Hemoglobin levels measured in POCTs were compared against the venous blood hemoglobin estimated in the hematology analyzer. Results A total of 120 children were enrolled. The mean (SD) of hemoglobin (g/dl) estimated in auto-analyzer was 9.4 (1.8), Sahli’s hemoglobinometer was 9.2 (1.9), invasive DH was 9.7 (1.9), and non-invasive DH was 11.9 (1.5). Sahli’s hemoglobinometer (95.5%) and invasive DH (92.2%) had high sensitivity for the diagnosis of anemia compared with non-invasive DH (24.4%). In contrast, non-invasive DH had higher specificity (96.7%) compared with invasive DH (83.3%) and Sahli’s hemoglobinometer (70%). Invasive DH took the least time (2–3 min) for estimation of hemoglobin per participant, followed by Sahli’s (4–5 min) and non-invasive DH (5–7 min). Conclusion All three POCT devices used in this study are reasonable and feasible for estimating hemoglobin in under-5 children. Invasive DHs are potential POCT devices for diagnosis of anemia among under-5 children, while Sahli’s can be considered as a possible option, where trained and skilled technicians are available. Further research and development are required in non-invasive DH to improve accuracy. Lay summary In India, anemia is a serious public health problem, where 58% of the children aged 6–59 months are anemic. Point-of-care testing (POCT) using digital hemoglobinometers (DHs) has been recommended as one of the key interventions by the Anemia Mukt Bharat program since 2018 in India. These POCT devices are easy to use, less invasive, can be carried to field, require minimal training and results are available immediately. Therefore this study assessed the validity of POCT devices—invasive DH, non-invasive DH and Sahli’s hemoglobinometer among 6–59 months children in facility setting compared with the gold standard hematology analyzer. A total of 120 children under 6–59 months of age were enrolled from the pediatric outpatient department in Haryana, India, from November 2019 to January 2020. The (mean hemoglobin in g/dl) invasive (9.7) and non-invasive DH (11.9) overestimated hemoglobin value, while Sahli’s (9.2) underestimated hemoglobin compared with hematology analyzer (9.4). Invasive DH (92.2%) and Sahli’s hemoglobinometer (95.5%) reported high ability to correctly identify those with anemia compared with non-invasive DH (24.4%). In contrast, non-invasive DH (96.73%) had higher ability to correctly identify those without the anemia compared with invasive DH (83.3%) and Sahli’s (70%).


Author(s):  
Chin-Pin Yeo ◽  
Carol Hui-Chen Tan ◽  
Edward Jacob

Background Point-of-care-testing (POCT) of haemoglobin Alc (HbA1c) is popular due to its fast turnaround of results in the outpatient setting. The aim of this project was to evaluate the performance of a new HbA1c POCT analyser, the Bio-Rad in2it, and compare it with the Siemens DCA 2000, Bio-Rad Variant II and Roche Tina-quant HbA1c Gen 2 assay on the cobas c501. Methods Imprecision of the four methods were compared by computing total imprecision from within-run and between-run data. A total of 80 samples were also compared and analysed by Deming regression and Altman–Bland difference test. Results Study of total imprecision of the in2it at HBA1c levels of 6.0% and 10.4% produced a coefficient of variation (%CV) of 3.8% and 3.7%, respectively. These results were more favourable as compared with the DCA 2000 but did not match the low imprecision of the central laboratory methods, the Bio-Rad Variant II and the Roche cobas c501. Comparison between the in2it and the central laboratory analysers, Bio-Rad variant II and cobas c501, revealed positive bias of 12% and 10%, respectively, supported by corresponding Deming regression equation slopes of +1.18 and +1.14. Comparison between the DCA 2000 and the central laboratory analysers revealed a bias that became increasingly positive with rising HbA1c concentrations with Deming regression analysis also revealing proportional and constant differences. Conclusions The in2it is a suitable POCT analyser for HbA1c but its less than ideal precision performance and differences with the central laboratory analysers must be communicated to and noted by the users.


Talanta ◽  
2020 ◽  
Vol 209 ◽  
pp. 120527 ◽  
Author(s):  
Ya-Jie Zhang ◽  
Shuai Chen ◽  
Yong-Liang Yu ◽  
Jian-Hua Wang

2018 ◽  
Vol 12 (3) ◽  
pp. 687-694 ◽  
Author(s):  
Karl-Heinz Patzer ◽  
Payam Ardjomand ◽  
Katharina Göhring ◽  
Guido Klempt ◽  
Andreas Patzelt ◽  
...  

Background: Medical practices face challenges of time and cost pressures with scarce resources. Point-of-care testing (POCT) has the potential to accelerate processes compared to central laboratory testing and can increase satisfaction of physicians, staff members, and patients. The objective of this study was to evaluate the effects of introducing HbA1c POCT in practices specialized in diabetes. Method: Three German practices that manage 400, 550, and 950 diabetes patients per year participated in this evaluation. The workflow and required time before and after POCT implementation (device: Alere Afinion AS100 Analyzer) was evaluated in each practice. Physician (n = 5), staff (n = 9), and patient (n = 298) satisfaction was assessed with questionnaires and interviews. Results: After POCT implementation the number of required visits scheduled was reduced by 80% (88% vs 17.6%, P < .0001), the number of venous blood collections by 75% (91% vs 23%, P < .0001). Of patients, 82% (vs 13% prior to POCT implementation) were able to discuss their HbA1c values with treating physicians immediately during their first visit ( P < .0001). In two of the practices the POCT process resulted in significant time savings of approximately 20 and 22 working days per 1000 patients per year (95% CI 2-46; 95% CI 10-44). All physicians indicated that POCT HbA1c implementation improved the practice workflow and all experienced a relief of burden for the office and the patients. All staff members indicated that they found the POCT measurement easy to perform and experienced a relief of burden. The majority (61.3%) of patients found the capillary blood collection more pleasant and 83% saw an advantage in the immediate availability of HbA1c results. Conclusions: The implementation of HbA1c POCT leads to an improved practice workflow and increases satisfaction of physicians, staff members and patients.


1996 ◽  
Vol 42 (5) ◽  
pp. 711-717 ◽  
Author(s):  
C A Parvin ◽  
S F Lo ◽  
S M Deuser ◽  
L G Weaver ◽  
L M Lewis ◽  
...  

Abstract We prospectively investigated whether routine use of a point-of-care testing (POCT) device by nonlaboratory operators in the emergency department (ED) for all patients requiring the available tests could shorten patient length of stay (LOS) in the ED. ED patient LOS, defined as the length of time between triage (initial patient interview) and discharge (released to home or admitted to hospital), was examined during a 5-week experimental period in which ED personnel used a hand-held POCT device to perform Na, K, Cl, glucose (Gluc), and blood urea nitrogen (BUN) testing. Preliminary data demonstrated acceptable accuracy of the hand-held device. Patient LOS distribution during the experimental period was compared with the LOS distribution during a 5-week control period before institution of the POCT device and with a 3-week control period after its use. Among nearly 15 000 ED patient visits during the study period, 4985 patients (2067 during the experimental period and 2918 during the two control periods) had at least one Na, K, Cl, BUN, or Gluc test ordered from the ED. However, no decrease in ED LOS was observed in the tested patients during the experimental period. Median LOS during the experimental period was 209 min vs 201 min for the combined control periods. Stratifying patients by presenting condition (chest pain, trauma, etc.), discharge/admit status, or presence/absence of other central laboratory tests did not reveal a decrease in patient LOS for any patient subgroup during the experimental period. From these observations, we consider it unlikely that routine use of a hand-held POCT device in a large ED such as ours is sufficient by itself to impact ED patient LOS.


Sign in / Sign up

Export Citation Format

Share Document